FDA Medical Device Recalls (openFDA)
Medline Industries, LP: Medline convenience kits: LOOP PACK DYNJ61460C OPEN HEART CDS840150Y PACEMAKER DYNJ56297C PACEMAKER PK DYNJT3454 PACEMAKER/ICD PACK DYNJ33822L PAD OPEN HEART DYNJ900962O PEDIATRIC OPEN HEART DYNJ54238L DYNJT6766 STAZ BASIC VASCULAR DYNJ909794C VASCULAR ACCESS KIT CVI5323A VASCULAR CDS CDS983040M
- Recall number
- Z-2418-2026
- Recalling firm
- Medline Industries, LP
- Product
- Medline convenience kits: LOOP PACK DYNJ61460C OPEN HEART CDS840150Y PACEMAKER DYNJ56297C PACEMAKER PK DYNJT3454 PACEMAKER/ICD PACK DYNJ33822L PAD OPEN HEART DYNJ900962O PEDIATRIC OPEN HEART DYNJ54238L DYNJT6766 STAZ BASIC VASCULAR DYNJ909794C VASCULAR ACCESS KIT CVI5323A VASCULAR CDS CDS983040M
- Status
- Open, Classified
- Initiated
- 2026-04-27
- Posted
- 2026-06-12
- Terminated
- not on file
- Reason
- Kits contain recalled lot of Ferndale Laboratories MASTISOL Liquid Adhesive. Butyrate tubing may crack during actuation and render product unusable for application.
- Root cause
- Process control
- Action
- Firm began notifying consignees on 4/27/2026. Customers were instructed to request stickers to over-label affected convenience kits with a warning label. Warning label indicates for user to remove the affected component and replace it with product from supply. No kits are to be returned.
- Quantity
- 49,654 kits total
- Distribution
- US Nationwide distribution. OUS distribution pending.
- Lot, serial or model codes
- DYNJ61460C UDI-DI 10198459422133 lot 25KBA098 CDS840150Y UDI-DI 10198459303029 lots 25JBW220 26BBD028 DYNJ56297C UDI-DI 10198459513954 lot 25LMJ874 DYNJT3454 UDI-DI 10198459325663 lot 25KBG761 DYNJ33822L UDI-DI 10198459385582 lot 26AMA072 DYNJ900962O UDI-DI 10198459452581 lot 26CBC302 DYNJ54238L UDI-DI 10198459443114 lot 25LMH549 DYNJT6766 UDI-DI 10198459580888 lot 25KBM870 DYNJ909794C UDI-DI 10198459528002 lot 25LBF702 CVI5323A UDI-DI 10198459582035 lot 25KBM758 CDS983040M UDI-DI 10198459579813 lots 25KBO682 26BBB926
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- OEZ
- Firm FEI
- 1417592
- Firm city
- Northfield
- Firm state
- IL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28