Skip to the content

FDA Medical Device Recalls (openFDA)

Securitas Healthcare LLC: Arial Call Station, Model Numbers CSK200-1069 and CSK200-1069MR, is a wireless nurse call station.

Recall number
Z-2418-2024
Recalling firm
Securitas Healthcare LLC
Product
Arial Call Station, Model Numbers CSK200-1069 and CSK200-1069MR, is a wireless nurse call station.
Status
Open, Classified
Initiated
2024-06-20
Posted
2024-07-23
Terminated
not on file
Reason
This is a risk that alarms will not be received by the Arial Wireless Nurse Call System and premature depletion of the battery.
Root cause
Software Manufacturing/Software Deployment
Action
Securitas Healthcare issued a recall notice on 06/19/2024 to its consignees via USPS certified mail. The notice explained the problem, potential hazard, and requested the following: "Customer Required Actions: 1. Closely Monitor Missing Device Alerts within the Arial Application and Immediately Verify Status of Missing Devices and Cease Using any Call Stations That Do Not Reconnect. Clients should closely monitor the "Missing Device Alerts within the Arial Application and immediately check the status of Call Stations that trigger a Missing Device Alert and replace the Call Station battery if it is low or fully depleted. Instructions for replacing the Call Station battery are attached hereto as Appendix 1. Instructions for reconnecting missing devices and confirming that the Call Stations are properly transmitting information to the Arial Emergency and Nurse Call System are attached hereto as Appendix 2. Clients should immediately cease using any Call Station that will not reconnect to the Arial Emergency and Nurse Call System. 2. Closely Monitor the Battery Status of all Call Stations in the Arial Web Client and Replace Battery if it is Low . Clients should closely monitor the Call Station Battery Status in the Arial Web Client and immediately replace the battery if it is low . 3. Consider Use of Supplementary Communication Devices or Procedures for High Risk Residents. Clients should consider enacting backup procedures, including increased room checks, or providing alternate emergency and nurse call communication means, such as a wearable pendant, for high-risk residents." "If you have distributed the product further, immediately notify your accounts that received the product identified above of this notification and ask them to contact Securitas Technical Support at 1-800-824-2996 (8:00 AM 5:00 PM CT, M-F) to obtain a response form and coordinate correction of affected devices." Securitas Healthcare is developing a firmware patch to remediate t
Quantity
77 units
Distribution
US: FL, CA, AL, ND, MS, WA, MN, TX, OR, WI, MD, NY, NE
Lot, serial or model codes
All units of the Affected Device Models sold between July 26, 2022 and June 5, 2024
510(k) numbers
not on file
PMA numbers
not on file
Product code
IQA
Firm FEI
1929691
Firm city
Lincoln
Firm state
NE
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register