EVIDENT SCIENTIFIC INC: SlideView VS200 slide viewer
- Recall number
- Z-2416-2024
- Recalling firm
- EVIDENT SCIENTIFIC INC
- Product
- SlideView VS200 slide viewer,
- Status
- Open, Classified
- Initiated
- 2024-07-22
- Posted
- 2024-07-26
- Terminated
- not on file
- Reason
- A broken fuse inside the laser that was caused by a short circuit at PIN 4 of the DOOR CTRL. This short circuit was caused by a badly positioned cable that made contact with the housing/frame and caused a short circuit and destroyed the fuse in the laser. This caused the shutter to remain in its most recent state; either in an open or closed position. Once the fuse was destroyed the shutter could not be operated.
- Root cause
- Radiation Control for Health and Safety Act
- Action
- not on file
- Quantity
- not on file
- Distribution
- US
- Lot, serial or model codes
- 65159 67105 67104 69153 67106 69154
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- RIB
- Firm FEI
- 3023822598
- Firm city
- Waltham
- Firm state
- MA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28