Skip to the content

FDA Medical Device Recalls (openFDA)

EVIDENT SCIENTIFIC INC: SlideView VS200 slide viewer

Recall number
Z-2416-2024
Recalling firm
EVIDENT SCIENTIFIC INC
Product
SlideView VS200 slide viewer,
Status
Open, Classified
Initiated
2024-07-22
Posted
2024-07-26
Terminated
not on file
Reason
A broken fuse inside the laser that was caused by a short circuit at PIN 4 of the DOOR CTRL. This short circuit was caused by a badly positioned cable that made contact with the housing/frame and caused a short circuit and destroyed the fuse in the laser. This caused the shutter to remain in its most recent state; either in an open or closed position. Once the fuse was destroyed the shutter could not be operated.
Root cause
Radiation Control for Health and Safety Act
Action
not on file
Quantity
not on file
Distribution
US
Lot, serial or model codes
65159 67105 67104 69153 67106 69154
510(k) numbers
not on file
PMA numbers
not on file
Product code
RIB
Firm FEI
3023822598
Firm city
Waltham
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.