Skip to the content

FDA Medical Device Recalls (openFDA)

Olympus Corporation of the Americas: ESG-410 Electrosurgical Generator- In conjunction with electrosurgical accessories intended for cutting and coagulation of tissue in open, laparoscopic , and endoscopic surgeries Model/Catalog Number: WA91307C

Recall number
Z-2415-2024
Recalling firm
Olympus Corporation of the Americas
Product
ESG-410 Electrosurgical Generator- In conjunction with electrosurgical accessories intended for cutting and coagulation of tissue in open, laparoscopic , and endoscopic surgeries Model/Catalog Number: WA91307C
Status
Open, Classified
Initiated
2024-06-21
Posted
2024-07-23
Terminated
not on file
Reason
Electrosurgical Generators ESG-410 does not power on upon use, may result in delays in initiating treatment or prolonged surgery
Root cause
Component design/selection
Action
Olympus issued URGENT: MEDICAL DEVICE REMOVAL Letter on 6/21/24. Letter states reason for recall, health risk and action to take: An Olympus representative will reach out to you to coordinate a repair to replace the power supply on your affected ESG-410 unit(s). Technical analysis of this issue performed by Olympus has determined that when the device is successfully powered on, it will work within specification. Therefore, you may continue to use your ESG-410 generator until the power supply is replaced. To minimize procedural delays, power on the generator prior to starting the procedure and, whenever possible, have a replacement device available. Additionally, Olympus requires you to take the following actions: 1. Carefully read the content of this notification. 2. If you have further distributed this product, identify your customers, and forward them this notification. 3. Acknowledge receipt of this letter through the Olympus web portal: a. Go to https://olympusamerica.com/recall Olympus requests that you report any complaints, including the ESG-410 generator not powering on, to the Technical Assistance Center (TAC) at 1- 800-848-9024 (option 1). If you require additional information, please do not hesitate to contact me by phone at (647) 999-3203 from Monday through Friday, 9 am to 5 pm EDT, or by e-mail at Cynthia.Ow@olympus.com.
Quantity
86
Distribution
US Nationwide distribution.
Lot, serial or model codes
UDI-DI: 04042761086349 Serial Numbers: 102410, 102409, 102408, 102419, 102418, 102417, 102416, 102420, 102412, 102407, 102394, 102393, 102401, 102400, 102405, 102404, 102403, 102428, 102424, 102422, 102438, 102432, 102431, 102430, 102429, 102437, 102435, 102433, 102442, 102440, 102439, 102436, 102366, 102364, 102372, 102371, 102444, 102441, 102502, 102499, 102498, 102497, 102515, 102513, 102512, 102511, 102506, 102504, 102510, 102505, 102503, 102501, 102500, 102522, 102520, 102518, 102517, 102516, 102523, 102519, 102396, 102369, 102434, 102392, 102823, 102822, 102821, 102819, 102818, 102802, 102814, 102810, 102809, 102808, 102834, 102830, 102829, 102828, 102919, 102925, 102917, 102916, 102913, 102912, 102836, 102921
510(k) numbers
["K203277"]
PMA numbers
not on file
Product code
GEI
Firm FEI
2429304
Firm city
Center Valley
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register