FDA Medical Device Recalls (openFDA)
Zimmer Inc.: NexGen complete knee solution MIS total knee procedure stemmed tibial component fixed bearing precoat size 8, sterile, REF 00-5950-057-02, Zimmer Inc. Warsaw, IN. This device is indicated for patients with severe knee pain and disability due to: Rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis. Collage disorders, and/or avascular necrosis of the femoral condyle. Post-traumatic loss of joint configuration, particularly when there is patellofemoral erosion, dysfunction or prior patellectomy. Moderate valgus, varus, or flexion deformities. The salvage of previously failed surgical attempts or for a knee in which satisfactory stability in flexion cannot be obtained at the time of surgery. NexGen TM Tibial Trays may be used with or without bone cement (biological fixation). NexGen MIS Tibial Components and MIS Modular Tibial Plates and Keels are intended for cemented use only.
- Recall number
- Z-2415-2010
- Recalling firm
- Zimmer Inc.
- Product
- NexGen complete knee solution MIS total knee procedure stemmed tibial component fixed bearing precoat size 8, sterile, REF 00-5950-057-02, Zimmer Inc. Warsaw, IN. This device is indicated for patients with severe knee pain and disability due to: Rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis. Collage disorders, and/or avascular necrosis of the femoral condyle. Post-traumatic loss of joint configuration, particularly when there is patellofemoral erosion, dysfunction or prior patellectomy. Moderate valgus, varus, or flexion deformities. The salvage of previously failed surgical attempts or for a knee in which satisfactory stability in flexion cannot be obtained at the time of surgery. NexGen TM Tibial Trays may be used with or without bone cement (biological fixation). NexGen MIS Tibial Components and MIS Modular Tibial Plates and Keels are intended for cemented use only.
- Status
- Terminated
- Initiated
- 2010-04-26
- Posted
- 2010-09-13
- Terminated
- 2011-03-28
- Reason
- The firm has received complaints of loosening of the implanted device requiring revision surgery. There have been 114 MDRs filed all reported that the device loosened and the patient required additional surgery to replace the device.
- Root cause
- Other
- Action
- The firm, Zimmer, Inc., sent an "URGENT DEVICE CORRECTION" letter dated April 26, 2010, to all customers. The letter described the product, problem and action to be taken by customers. The customers were instructed to destroy or disregard all previous versions of the surgical technique, review the attached revised surgical technique to familiarize yourself with the modifications and warnings, and complete and return the Health Care Professional Dcoumentation Form via fax to Zimmer, Inc. at (574) 372-4265. (Note: Zimmer made a modification to the surgical technique and instructions for use. They added a warning to fully cement and pressurize the anterior and posterior surfaces of the tibial component, and to strongly recommend the use of a drop down stem extension. This change was completed and approved on April 23, 2010.) For questions please contact Zimmer at 1-877-946-2761.
- Quantity
- 68,384 all products
- Distribution
- Worldwide distribution: USA including states of AZ, IL, IN, KY, MO, NE, NM, NY, OH, TX, VA, WA, and WI; and countries of Austria, Belgium, France, Germany, Italy, Netherlands, Spain, Sweden, Switzerland, UK, Ausralia, China, Hong Kong, India, Japan, Korea, Malaysia, Nicaragua, Singapore, Tawain and Thailand.
- Lot, serial or model codes
- 103044, 60276808, 60282368, 60292625, 60296123, 60296124, 60307227, 60315651, 60328159, 60331388, 60331389, 60339651, 60343929, 60343930, 60343931, 60343932, 60343933, 60343934, 60343935, 60355713, 60362083, 60365658, 60384329, 60534175, 60570899, 60578778, 60584754, 60616292, 60625488, 60640829, 60692575, 60727493, 60760516, 60818767, 60867946, 60892877, 60917518, 60955232, 60983707, 61113641 and 61143803
- 510(k) numbers
- ["K043101"]
- PMA numbers
- not on file
- Product code
- JWH
- Firm FEI
- 1000220733
- Firm city
- Warsaw
- Firm state
- IN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28