Skip to the content

FDA Medical Device Recalls (openFDA)

Palodex Group Oy: Orthopantomograph OP300 dental computed tomography x-ray system. The X-ray device produces conventional 2D X-ray images and Xray projection images for the reconstruction of a 3D view. The device is operated and used by qualified healthcare professionals.

Recall number
Z-2413-2024
Recalling firm
Palodex Group Oy
Product
Orthopantomograph OP300 dental computed tomography x-ray system. The X-ray device produces conventional 2D X-ray images and Xray projection images for the reconstruction of a 3D view. The device is operated and used by qualified healthcare professionals.
Status
Open, Classified
Initiated
2021-10-28
Posted
2024-08-12
Terminated
not on file
Reason
It was noticed in production on 28.10.2021, that the final testing database gives a pass result for Pan ADC repeatability test even when the test result is not within the set acceptance criteria (repeatability within 4%).
Root cause
Radiation Control for Health and Safety Act
Action
Palodex Group Oy is preparing a Class IlIl field correction to update the firmware of ORTHOPANTOMOGRAPH OP 3D Pro of the 27 defective units and these devices will be retested for Pan ADC reproducibility. All affected persons will be notified of this defect in accordance with 21 CFR 806.10.
Quantity
103
Distribution
US Nationwide-Worldwide
Lot, serial or model codes
DI: 06430035872725, 06430035872732, 06430035872749, 06430035872756, 06430035872763, 06430035872770
510(k) numbers
["K163423"]
PMA numbers
not on file
Product code
OAS
Firm FEI
3005383085
Firm city
Tuusula
Firm state
not on file
Firm country
Finland

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register