Skip to the content

FDA Medical Device Recalls (openFDA)

Waldemar Link GmbH & Co. KG (Mfg Site): Endo Model SL Tibial Components 16-2817/02 Tibial Component, Small, W: 60mm 16-2817/05 Tibial Component, Medium, W: 65mm 16-2817/07 Tibial Component, Large, W: 75mm 16-2817/32 Tibial Component Modular, Small, W: 60mm 16-2817/35 Tibial Component Modular, Medium, W: 65mm 16-2817/37 Tibial Component Modular, Large, W: 75mm

Recall number
Z-2411-2024
Recalling firm
Waldemar Link GmbH & Co. KG (Mfg Site)
Product
Endo Model SL Tibial Components 16-2817/02 Tibial Component, Small, W: 60mm 16-2817/05 Tibial Component, Medium, W: 65mm 16-2817/07 Tibial Component, Large, W: 75mm 16-2817/32 Tibial Component Modular, Small, W: 60mm 16-2817/35 Tibial Component Modular, Medium, W: 65mm 16-2817/37 Tibial Component Modular, Large, W: 75mm
Status
Open, Classified
Initiated
2024-05-22
Posted
2024-07-18
Terminated
not on file
Reason
Blind screws of the modular tibial component cannot be loosened intraoperatively,prolongation surgery due to intraoperatively change in procedure, probably to cementing technique
Root cause
Process design
Action
LinkBio issue Urgent Medical Device Correction letter to Distributors via email on June 10, 2024. Letter states reason for recall, health risk and action to take: Please take note of the attached modified Surgical Technique for cementing the Tilastan Proximal Tibial Spacer or Proximal Tibial Segment, provided in Appendix A of this notice. No product return is required. Please respond using the attached Distributor Reply form to indicate your understanding of this Device Correction and of the need to apply the provided cementing surgical technique for the Tilastan Proximal Tibial Spacer or Proximal Tibial Segment, in case the blind screws are unable to be removed. Please contact LinkBio at 973-625-1333 or complaints@linkbio.com for any questions
Quantity
114 units US; 2351 units OUS
Distribution
Nationwide Foreign: Algeria Argentina Australia Austria Belarus Belgium Brazil Bulgaria Canada China Colombia Czechia Denmark Ecuador Estonia Finland France Germany Greece Hungary India Indonesia Israel Italy Kenya Lithuania Luxembourg Mexico Netherlands Norway Peru Poland Romania Saudi Arabia Slovakia Slovenia Spain Sweden Switzerland Thailand T¿rkiye Ukraine United Arab Emirates United Kingdom Uruguay Vietnam
Lot, serial or model codes
Product UDI-DI 16-2817/02 04026575359202; 16-2817/05 04026575359219; 16-2817/07 04026575359226; 16-2817/32 04026575370870; 16-2817/35 04026575370887; 16-2817/37 04026575370894. All product lots manufactured since 01-Jun-2022.
510(k) numbers
["K212742","K143179","K151008"]
PMA numbers
not on file
Product code
JWH
Firm FEI
3003386935
Firm city
Norderstedt
Firm state
not on file
Firm country
Germany

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register