Elekta, Inc.: Monaco Radiation Treatment Planning System (RTP) System

Recall number
Z-2409-2019
Recalling firm
Elekta, Inc.
Product
Monaco Radiation Treatment Planning System (RTP) System
Status
Terminated
Initiated
2019-08-23
Posted
not on file
Terminated
2024-06-28
Reason
Monaco is displaying the Anatomy and Beam shift direction incorrectly on the Monaco Scan and Setup Reference Report and is DICOM exporting incorrect (Anatomy/Beam) shift directions.
Root cause
Device Design
Action
Elekta Inc. notified customers on about 08/23/2019 via "IMPORTANT FIELD SAFETY NOTIFICATION" letter. The letter informed customers that the Monaco is displaying some incorrect shift information on the Scan and Setup Reference Report and is DICOM exporting incorrect shift information. Instructions included that all shifts should be verified before treatment, that a software fix is planned. Customer were also instructed to post the notice in a place accessible to all users, advise personnel working with the device of the issue, and to complete and return the Acknowledgement Form provided.
Quantity
49 units
Distribution
Distribution within the US to Ohio and Indiana. International distribution to Germany, Greece, Sweden.
Lot, serial or model codes
Software Build: 5.50.00 & 5.51.00; UDI GTIN: (01)00858164002244(10) 5.50.00, (01)00858164002268(10) 5.51.00
510(k) numbers
K190178
PMA numbers
not on file
Product code
MUJ
Firm FEI
1037831
Firm city
Atlanta
Firm state
GA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this source