Skip to the content

FDA Medical Device Recalls (openFDA)

Alcon Research, LTD.: Constellation Procedure Pak

Recall number
Z-2406-2023
Recalling firm
Alcon Research, LTD.
Product
Constellation Procedure Pak
Status
Open, Classified
Initiated
2023-06-28
Posted
2023-08-16
Terminated
not on file
Reason
Reports of patient injury during use in procedures due to potential for plastic debris from over-tightening with included plastic wrench.
Root cause
Process control
Action
Customers were sent a Voluntary Medical Device Correction Notification by mail and dated 06/29/2023. Customers are asked to follow the risk mitigation precautions detailed in the notice to avoid the instance of overtightening the Phaco Tip and generating plastic particles. Additionally, customers are asked to forward the recall notification to all departments or organizations that use recalled devices. Customers are asked to complete and return the provided Response Form by fax at 817-302-4337 or email at Market.Actions@alcon.com. Customers who have experienced any adverse events or quality related issues are to contact alcon via email at msus.safety@alcon.com or by phone at 1-800-757-9780, Option 5.
Quantity
10,580 units
Distribution
Worldwide distribution - US Nationwide and the countries of Aland Islands, Albania, Algeria, Andorran, Argentina, Armenia, Aruba, Australia, Austria, Azerbaijan, Bahamas, Bahrain, Bangladesh, Barbados, Belarus, Belgium, Bermuda, Botswana, Brazil, Brunei Daruss., Bulgaria, Burma, Cambodia, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, Croatia, Cura¿ao, Cyprus, Czech Republic, Denmark, Dominican Rep., Ecuador, Egypt, El Salvador, Finland, France, French Polynesia, French Guayana, Georgia, Germany, Ghana, Greece, Guadeloupe, Guernsey, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Isle of Man, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Kosovo, Kuwait, Kyrgyzstan, Latvia, Lebanon, Libya, Luxembourg, Macau, Malaysia, Malta, Martinique, Mauritius, Mexico, Moldova, Mongolia, Montenegro, Morocco, Namibia, Netherlands, New Caledonia, New Zealand, Nicaragua, North Macedonia, Norway, Oman, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, Reunion, Romania, Russian Fed., Saint Martin, San Marino, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Syria, Taiwan, Thailand, Trinidad, Tobago, Tunisia, Turkey, Ukraine, United Kingdom, Uruguay, USA, USVI, Utd. Arab Emir., Uzbekistan, Vietnam, and Yemen.
Lot, serial or model codes
Catalog No. 8065750958; Product Code 100007216; UDI-DI: 380657509584; Lots 14XM2J, 14Y1H6, 150HYJ, 150HYK, 151R72, 152CH0, 15309X, 156HWN, 15A13R, 15A13T, 15CF2X, 15CF2Y, 15CF30, 15CJ47, 15CJ48, 15CWAM, 15CWAN, 15D649.
510(k) numbers
["K141065"]
PMA numbers
not on file
Product code
HQC
Firm FEI
2523835
Firm city
Sinking Spring
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register