Skip to the content

FDA Medical Device Recalls (openFDA)

Acist Medical Systems: ACIST" BRACCO, AT P54 AngioTouch Kit , SKU # 014644, Sterile EO Manufacturer; ACTB Medical Systems, Inc. 7905 Fuller Road, Eden Prairie, MN 55344, Made in Mexico. Catalogue No: 800608-013 Intended to be used for the controlled infusion of radiopaque contrast media for angiographic procedures.

Recall number
Z-2406-2008
Recalling firm
Acist Medical Systems
Product
ACIST" BRACCO, AT P54 AngioTouch Kit , SKU # 014644, Sterile EO Manufacturer; ACTB Medical Systems, Inc. 7905 Fuller Road, Eden Prairie, MN 55344, Made in Mexico. Catalogue No: 800608-013 Intended to be used for the controlled infusion of radiopaque contrast media for angiographic procedures.
Status
Terminated
Initiated
2008-06-06
Posted
2008-09-23
Terminated
2011-01-25
Reason
Some of the sterile package seals of AT P54 AngioTouch Kit were breached. Breached package seals could compromise product sterility, which could potentially lead to patient infection.
Root cause
Package design/selection
Action
Consignees were sent a "Urgent: Medical Device Recall" letter dated July 9. 2008. The letter addressed to "Manager" described the Proben and product. The letter advised customers that if they had inventory of any of the lots under the scope of the recall, to stop further distribution of that product and immediately cease use of the product. The letter listed the applicable lots and asked customers to verify their inventory against the list. Any inventories matching the lots on the recall list were to be checked on the provided form and the quantity recorded. They were instructed to return the product to ACIST using the provided Air Bill number along with the form, signed and dated. The letters were sent July 10, 2008.
Quantity
11,750 (USA), 22,230 (OUS)
Distribution
Worldwide Distribution --- USA including states AR, AZ, DC, FL, GA, MS, NC, NE, NH, NY, OH, OK, PA, TX, TN,VA, AL, CA, IL, KS, KY, LA, MD, MN, MO, SD, CO, IA, IN, MI, NV, PR, SC, MI, OR, WV, CT, ND, DC, WA, ME, and NJ, and countries of Canada, Brazil, Japan, China, Korea, Thailand, Hong Kong, and Australia.
Lot, serial or model codes
0368H, 0368J, 0378L, 0388F, 0388N, 0388V, 0498F, 0498H, 0508B, 0528P, 0528W, 0578N, 0588J, 0588K, 0598T, 0848K, 0848L, 0878M, and 0918N.
510(k) numbers
["K040298"]
PMA numbers
not on file
Product code
DXT
Firm FEI
3001726808
Firm city
Eden Prairie
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register