Cordis Corporation: RenLane Renal Denervation Catheter; Cat No.: D135601. Used in adults patients (>18 years) with drug resistant hypertension to denervate the renal arteries to reduce blood pressure.

Recall number
Z-2405-2015
Recalling firm
Cordis Corporation
Product
RenLane Renal Denervation Catheter; Cat No.: D135601. Used in adults patients (>18 years) with drug resistant hypertension to denervate the renal arteries to reduce blood pressure.
Status
Terminated
Initiated
2014-04-15
Posted
2015-08-18
Terminated
2015-08-26
Reason
Potential Damage may occur to the helical tip of the RENLANE Renal Denervation Catheter, if the hemostasis valve of the arterial access device is not fully open before insertion or withdrawal of the catheter.
Root cause
Labeling Change Control
Action
Cordis sent an Urgent Field Safety Notice dated April 15, 2014 providing additional labeling to their consignees in Germany only, with a follow-up letter on August 4, 2014. Cordis explained When deploying the catheter via a compatible arterial access device, if a Tuohy-Borst type adjustable hemostasis valve is used, ensure the valve is fully open before insertion or withdrawal of the catheter. This is especially important, since the helical tip at the distal end of the RENLANE Renal Denervation Catheter has a larger diameter compared to the proximal catheter shaft. Although the event was not a Removal, all lots involved have expired. The product is no longer commercialized, and was never commercialized beyond the 10 customers in Germany.
Quantity
98 units
Distribution
Distributed in Germany only.
Lot, serial or model codes
Model Number: D135601 Lots #'s 15958389LA, 15959088LA, 15959090LA, 15968753LA, 15968806LA, 15968810LA, and 15968813LA, with Exp Date 8/31/2014 Lots #'s 15976735L, 15976820L, 15976823L, 15978174L, 15978179L, 159784180L, and 15978181L with Exp Date 9/30/2014
510(k) numbers
not on file
PMA numbers
not on file
Product code
MPB
Firm FEI
1016427
Firm city
Miami Lakes
Firm state
FL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register