Philips Medical Systems North America Co. Phillips: EasyDiagnost Classic has a radiographic tilting table. This system has a cassette film option and it does not have a Digital Spot Imaging (DSI) system. Intended use is for the following applications: as a multifunctional/universal system, general R/F, Fluoroscopy, Radiology, and Angiography can be performed along with pediatric examinations and more specialized interventional applications.

Recall number
Z-2405-2010
Recalling firm
Philips Medical Systems North America Co. Phillips
Product
EasyDiagnost Classic has a radiographic tilting table. This system has a cassette film option and it does not have a Digital Spot Imaging (DSI) system. Intended use is for the following applications: as a multifunctional/universal system, general R/F, Fluoroscopy, Radiology, and Angiography can be performed along with pediatric examinations and more specialized interventional applications.
Status
Terminated
Initiated
2008-01-04
Posted
2010-09-13
Terminated
2010-09-13
Reason
Potential for the tower (holding the serial changer and/or image intensifier) to slide unexpectedly and potentially injure a patient or user in its path.
Root cause
Other
Action
On 1/4/08, the firm issued the URGENT PRODUCT CORRECTION letter dated 1/3/08 to their consignees via certified mail, return receipt requested. The letter informs consignees the issue of the safety switch and counter weight steel band on the Diagnost 55/66/76Plus, Easy Diagnost, and Easy Diagnost Eleva X-ray systems. However, the Easy Diagnost Eleva X-ray system was distributed and applied only to foreign consignees outside US. If the consignees hear the primary steel band clicking/rattling noise while tilting the table to 90 degrees table or positioning the serial changer, the consignees are instructed to tilt the table base towards the 0 degree tilting position and do not use the tilting drive movement until the mandatory action by a Philips service representative has been performed. The Field Change Order (FCO) would be implemented on 01/23/08 and expected to be completed by July 2008. The firm's representatives will visit sites and will inspect all systems where the switch has been replaced or adjusted for proper operation of the safety sensor and switch. Two steel bands will be inspected and replaced if any abnormity is noted. In the mean time, consignees should contact Philips Call Center at 800-722-9377, #5, #2 and reference the FCO 70600029 with any questions.
Quantity
1 unit
Distribution
Nationwide Distribution
Lot, serial or model codes
Device is identified as Site number 86389.
510(k) numbers
["K970640"]
PMA numbers
not on file
Product code
IXR
Firm FEI
1000524572
Firm city
Bothell
Firm state
WA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register