FDA Medical Device Recalls (openFDA)
Customed, Inc: RETINA PACK - (1) TABLE COVER REINFORCED 44" X 78" LIF (50) COTTON TIP APPLICATOR 6" WOOD LIF (2) GOWN IMPERVIOUS EXTRA REINFORCED LARGE (2) MEDECINE CUP 2oz LIF (5) TOWELS ABSORBENT 15" X 20" LIF ( 1) UNIVERSAL EYE SHIELD (1) EYE DRAPE (1) BIPOLAR CABLE DISPOSABLE ( 1) OPHTALMIC BIPOLAR PENCIL (10) GAUZE SPONGE 4" X 4" 16PLY (2) EYE PAD 2 1/8 X 2 5/8 ST. ( 1) MAYO STAND COVER REINFORCED LIF ( 1) KNIFE 15 5.0 CONTROLLED DEPTH. (1) NEEDLE & BLADE COUNTER 20C FOAM/MAG LIF (1) SURGICAL BLADE #15 STAINLESS STEEL (3) SYRINGE 1CC TUBER. 25G X 5/8 SLIP TIP L/F EO sterilized surgical convenient packs packaged into polyester/polyethylene breather bag with Tyvek double vents.
- Recall number
- Z-2403-2014
- Recalling firm
- Customed, Inc
- Product
- RETINA PACK - (1) TABLE COVER REINFORCED 44" X 78" LIF (50) COTTON TIP APPLICATOR 6" WOOD LIF (2) GOWN IMPERVIOUS EXTRA REINFORCED LARGE (2) MEDECINE CUP 2oz LIF (5) TOWELS ABSORBENT 15" X 20" LIF ( 1) UNIVERSAL EYE SHIELD (1) EYE DRAPE (1) BIPOLAR CABLE DISPOSABLE ( 1) OPHTALMIC BIPOLAR PENCIL (10) GAUZE SPONGE 4" X 4" 16PLY (2) EYE PAD 2 1/8 X 2 5/8 ST. ( 1) MAYO STAND COVER REINFORCED LIF ( 1) KNIFE 15 5.0 CONTROLLED DEPTH. (1) NEEDLE & BLADE COUNTER 20C FOAM/MAG LIF (1) SURGICAL BLADE #15 STAINLESS STEEL (3) SYRINGE 1CC TUBER. 25G X 5/8 SLIP TIP L/F EO sterilized surgical convenient packs packaged into polyester/polyethylene breather bag with Tyvek double vents.
- Status
- Terminated
- Initiated
- 2014-05-20
- Posted
- 2014-08-29
- Terminated
- 2017-08-14
- Reason
- Customed has determined that there is the possibility that packaging integrity may be compromised on different catalog numbers (multiple lots). This could result in an injury to the patient due to product contamination or loss of sterility condition.
- Root cause
- Package design/selection
- Action
- Customed Inc. sent a Recall Notification letter dated June 3, 2014 via email and followed by letter to all affected customers. The letter identified the affected products, problem and actions to be taken. The letter informed customers if the affected products are in their inventory to: 1. Remove products from sale (quarantine them) and cease distribution. 2. Complete and return the enclosed acknowledgement form to Customed Inc .Recall Coordinator. Affected products should be returned to Customed Inc for credit. Customers with questions were instructed to contact Customed Inc. Recall Coordinator at 1-787-801-0100 Ext. 7540
- Quantity
- 10 lots; 336 units (multiple units per lot)
- Distribution
- Worldwide Distribution - US (nationwide) in the state of Florida including Puerto Rico (USA); & USVI (St John) and the countries of Tortola, Trinidad, Dominican Republic, Costa Rica and Mexico.
- Lot, serial or model codes
- Product code 900-1611, 10 lots: 111123361 112030624 112051849 112072779 112114640 112124965 113078268 113088833 140412455 140513000
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- OJK
- Firm FEI
- 2648727
- Firm city
- Fajardo
- Firm state
- not on file
- Firm country
- PR
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as