Skip to the content

FDA Medical Device Recalls (openFDA)

Steris Corporation: A mobile, electro-hydraulically operated, fluoroscopic/C-Arm compatible general surgical table. Product Usage: Product provides a flexible, easy-to-use, articulated posturing of a patient for either general, I.A., or specialized surgical procedures. There are movable components and accessories intended for use during diagnostic examinations or surgical procedures to support and position a patient.

Recall number
Z-2402-2012
Recalling firm
Steris Corporation
Product
A mobile, electro-hydraulically operated, fluoroscopic/C-Arm compatible general surgical table. Product Usage: Product provides a flexible, easy-to-use, articulated posturing of a patient for either general, I.A., or specialized surgical procedures. There are movable components and accessories intended for use during diagnostic examinations or surgical procedures to support and position a patient.
Status
Terminated
Initiated
2010-02-17
Posted
2012-09-17
Terminated
2013-03-13
Reason
Customers may be storing objects on the base and/or around the column of the surgical table. This prohibited practice and misuse known by the customers can damage the override switch assembly located at the top of the column cover. Depending upon the nature and severity of the damage to the electronics of the table, it is possible that the table could move without prompting by an operator .
Root cause
Employee error
Action
Steris sent an Urgent Field Correction Notice letter dated February 17, 2010 to their customers. The letter identified the affected product, description of the problem and actions to be taken. Customers were instructed to examine their units for any structural damage. If found they are to discontinue use and arrange for a field representative to visit the facility to assess the damages and arrange for repair of the unit. For information or questions regarding STERIS visit to your facility contact STERIS Field Service Dispatch at 1-800-333-8828.
Quantity
11,044 Units
Distribution
Worldwide Distribution - US Nationwide including District of Columbia and Puerto Rico and the following countries: American Samoa, Argentina, Australia, Bahrain, Bangladesh, Bermuda, Brazil, Canada, China, Colombia, Costa Rica, Dominican Republic, Ecuador, Egypt, Germany, Greece, Guam, Hong Kong, India, Indonesia, Israel, Italy, Japa, Jordan, Kazakhstan, Korea, Kuwait, Lebanon , Libya , Malaysia , Mexico, Morocco, New Zealand, Panama, Peru, Philippines , Poland, Portugal, Qatar, Russia, Saudi Arabia , Singapore, South Africa, Spain, Sri Lanka, Taiwan, Thailand, Trinidad and Tobago, Turkey & United Kingdom.
Lot, serial or model codes
B420702032 to 0402110097
510(k) numbers
not on file
PMA numbers
not on file
Product code
FQO
Firm FEI
1527821
Firm city
Mentor
Firm state
OH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register