Skip to the content

FDA Medical Device Recalls (openFDA)

Cepheid: Xpert Infinity Software, V 4.1a. Cepheid, Sunnyvale, CA 94089 system software for GeneXpert Infinity 48

Recall number
Z-2401-2011
Recalling firm
Cepheid
Product
Xpert Infinity Software, V 4.1a. Cepheid, Sunnyvale, CA 94089 system software for GeneXpert Infinity 48
Status
Terminated
Initiated
2011-04-01
Posted
2011-05-31
Terminated
2011-06-20
Reason
A mismatch of Patient ID or Sample ID and the Host Order from the hosted LIS and its requested test results could occur when running in batch mode and cartridges are introduced in rapid succession.
Root cause
Software design
Action
Cepheid sent an Corrective Action Notification letter dated April 13, 2011, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Cepheid recommended customers using 4.1a software on the INFINITY 48 System that is interfaced to a LIS with Host Ordering to make these alterations in operationof the system. After scanning the Patient ID (optional). Sample ID and Cartridge barcode, the host test order information should apear highlighted in the Order Test workspace. Once the highlighted host order information appears select SUBMIT. Warning. If the host test order informaiton is not highlighted, cancel the order by selecting CLOSE. Re scan Patient ID (optional). Sample ID and cartridge barcode and verify highlighted host test order before SUBMIT. Contact Cepheid Technical Support at 888-838-3222 in the US or 33 5 63 82 53 19 outside of the US for any questions regarding this Corrective Action Notice.
Quantity
16 US, 3 ROW
Distribution
Worldwide Distribution - USA including CA, UT, TX, NJ, FL, NC, CO, IN, VT, MN and the countries of the UK, Denmark and Saudi Arabia.
Lot, serial or model codes
V 4.1a
510(k) numbers
["K103766"]
PMA numbers
not on file
Product code
NJR
Firm FEI
3004530258
Firm city
Sunnyvale
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register