Angiodynamics, Inc.: Centros(TM) 15F x 28cm Chronic Hemodialysis Curved Catheter Set (With Cuff 23cm From Tip), Catalog No./REF 10303102. Sterile. Manufactured for AngioDynamics, Inc. MANUFACTURED IN USA. ANGIODYNAMICS, INC. Queensbury, NY USA 12804.

Recall number
Z-2397-2008
Recalling firm
Angiodynamics, Inc.
Product
Centros(TM) 15F x 28cm Chronic Hemodialysis Curved Catheter Set (With Cuff 23cm From Tip), Catalog No./REF 10303102. Sterile. Manufactured for AngioDynamics, Inc. MANUFACTURED IN USA. ANGIODYNAMICS, INC. Queensbury, NY USA 12804.
Status
Terminated
Initiated
2008-07-07
Posted
2008-09-24
Terminated
2012-08-02
Reason
The cuff may be inadequately attached to the catheter, resulting in possible catheter movement or leakage at the insertion site.
Root cause
Process design
Action
AngioDynamics sent URGENT MEDICAL DEVICE letters and reply forms to the consignees via Certified Mail on/about 7/10/08. The recall is being conducted to the medical user level. AngioDynamics is requesting the return of all Centros Hemodialysis Catheters in inventory. The following steps should be taken by the consignees: 1. Identify and segregate the recalled product that is in your possession. 2. Complete the enclosed Centros Hemodialysis Catheter Recall Reply Form and fax it to the attention of the Centros Recall Coordinator at 518-798-1360. It is important that even if you do not have any product remaining in your possession that you fill out the attached form noting zero quantity to be returned and fax the form to AngioDynamics. 3. Ship the recalled product(s) to AngioDynamics. Consignees will be credited for the returned products. Credits will only be issued for returned product that is within the scope of the recall. Reference Return Authorization Number ___ on the outside of the shipping box and include a copy of the Centros Recall Reply Form with the return shipment. --- The letters also provide actions to be taken for product(s) in use: ANGIODYNAMICS recommends that physicians continue to follow facility protocols for inserting dialysis catheters and monitoring dialysis catheter patients. It is recommended that physicians consider evaluation of patients with an indwelling Centros catheter to assure the catheter is properly secured. As stated in the product's instructions for use, it is recommended that the catheter be secured/sutured for the entire duration of implantation. If the sutures have been cut, consider re-suturing to maintain proper catheter position. Do not rely on ingrowth of the cuff to maintain proper catheter position. This failure does not affect the function or integrity of the catheter itself and there have no reports of patient injury due to this problem. For questions, concerns, or more detailed instructions on how to
Quantity
not on file
Distribution
Nationwide Distribution
Lot, serial or model codes
Lots S08B02, S08B06, S08C01, S08D02, and S08D11
510(k) numbers
["K070572"]
PMA numbers
not on file
Product code
MSD
Firm FEI
1319211
Firm city
Queensbury
Firm state
NY
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register