Ad-Tech Medical Instrument Corporation: AD-TECH Spencer Probe Depth Electrode Product Usage: The Ad-Tech Depth Electrodes (Depth Electrodes, Foramen Ovale Depth Electrodes, Macro Micro Depth Electrodes, Spencer Probe Depth Electrodes, Wyler Sphenoidal Depth Electrodes) are intended for temporary (< 30 days) use with recording, monitoring and stimulation equipment for the recording, monitoring and stimulation of electrical signals at the subsurface level of the brain. The recording of electrical activity supports definition of the location of epileptogenic foci and brain mapping.

Recall number
Z-2396-2019
Recalling firm
Ad-Tech Medical Instrument Corporation
Product
AD-TECH Spencer Probe Depth Electrode Product Usage: The Ad-Tech Depth Electrodes (Depth Electrodes, Foramen Ovale Depth Electrodes, Macro Micro Depth Electrodes, Spencer Probe Depth Electrodes, Wyler Sphenoidal Depth Electrodes) are intended for temporary (< 30 days) use with recording, monitoring and stimulation equipment for the recording, monitoring and stimulation of electrical signals at the subsurface level of the brain. The recording of electrical activity supports definition of the location of epileptogenic foci and brain mapping.
Status
Terminated
Initiated
2019-04-16
Posted
not on file
Terminated
2022-05-27
Reason
An Ad-Tech Clinical Specialist, attended a case on April 15, 2019. During the case the surgeon encountered an issue after implanting four Ad-Tech depth electrodes. Upon initial device testing subsequent to implantation it was identified that two of the devices that were labeled as 8 contact depth electrodes (SD08R-SP05X-000) were in fact 6 contact depth electrodes (SD06R-SP05X-000) . There was no impact to the patient due to this incident.
Root cause
Labeling Change Control
Action
Via a MEDICAL DEVICVE RECALL letter dated, April 17, 2019, Ad-Tech advised their consignees of the mix-up, requested that they check their stock, returned the MEDICAL DEVICE RECALL RETURN RESPONSE, and return the subject devices. Distributors were requested to conduct a sub-recall to their customers.
Quantity
18
Distribution
Worldwide - Nationwide distribution in the state of Illinois and countries of Russia and United Kingdom
Lot, serial or model codes
Catalog Numbers SD08R-SP05X-000 SD06R-SP05X-000  LOT NUMBERS 127219 208140699 127357 208140699
510(k) numbers
["K163355"]
PMA numbers
not on file
Product code
GZL
Firm FEI
2183456
Firm city
Oak Creek
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register