Sedecal S.A.: RadPro 40kW Digital X-Ray Panel and Mobile/Portable Diagnostic X-Ray. Intended for use by a qualified/trained doctor or technician on both adult and pediatric subjects for taking diagnostic radiographic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts.

Recall number
Z-2390-2015
Recalling firm
Sedecal S.A.
Product
RadPro 40kW Digital X-Ray Panel and Mobile/Portable Diagnostic X-Ray. Intended for use by a qualified/trained doctor or technician on both adult and pediatric subjects for taking diagnostic radiographic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts.
Status
Terminated
Initiated
2015-07-13
Posted
2015-08-14
Terminated
2018-02-20
Reason
Electrostatic energy may be stored in Systems that can cause a short-cicuit which, in turn, can generate an Electomagnetic Pulse which might disturb the electronics of the Digital Motion Control Board producing uncontrolled movements and such uncontrolled movements could cause the system to bump into the user or ppatient.
Root cause
Device Design
Action
FIELD CORRECTION NOTIFICATION letters dated July 13, 2015 were sent to all dealers. The letters instructed the dealers: to schedule the replacement of the current arrestor by a new design of resistive arrestor with a resistance of 1 M ohm between frame and floor (ground). Although the metallic arrestor is only implemented in the equipments manufactured from April 2013 till March 2015, the upgrade kit has to be implemented in all the installed base of Mobiles with the same configuration in order to ensure that even the arrestors installed as spare part are replaced. Sedecal will provide the necessary upgrade kits to all the customers or distributors. *** U.S. Distributor, Virtual Imaging (a Canon U.S.A. Company) will use letter entitled, FIELD CORRECTION NOTIFICATION, RADPRO 40kW - -DIGITAL MOBILE X-RAY SYSTEM and dated 7/13/2015 to communicate to end users. The letter states "Once Virtual Imaging receives the resistive arrestor, we will contact you to schedule the replacement on the Affected System that you service." Please contact Virtual Imaging, Inc.'s Technical Support for any questions at 561.893.8500 or technicalsupport@vifla.com. ****
Quantity
361 systems (U.S.); 5 systems (Foreign)
Distribution
Distributed US (nationwide) and in Canada.
Lot, serial or model codes
Serial numbers: G29393 , G29395 , G29404 , G29416 , G29426 , G29864 , G29873 , G29878 , G30025 , G30155 , G30156 , G30160 , G30165 , G30173 , G30236 , G30240 , G30250 , G30254 , G30258 , G30567 , G30570 , G30684 , G30689 , G30761 , G30762 , G30771 , G30808 , G30902 , G30952 , G30960 , G31065 , G31077 , G31252 , G31256 , G31259 , G31271 , G31285 , G31403 , G31412 , G31526 , G31761 , G32487 , G32498 , G32514 , G32577 , G32581 , G32584 , G32587 , G32657 , G32671 , G32753 , G32936 , G33004 , G33102 , G33103 , G33111 , G33112 , G33126 , G33127 , G33144 , G33212 , G33238 , G33309 , G33324 , G33333 , G33344 , G33581 , G33585 , G33634 , G33708 , G33711 , G33721 , G33723 , G35389 , G35422 , G35429 , G35430 , G35442 , G35450 , G35462 , G35480 , G35504 , G35524 , G35611 , G35620 , G35630 , G35641 , G35665 , G35718 , G36042 , G36137 , G36148 , G36206 , G36211 , G36217 , G36222 , G36227 , G36244 , G36272 , G36281 , G36287 , G36468 , G36492 , G36526 , G36537 , G36542 , G36566 , G36573 , G36581 , G36601 , G36626 , G36643 , G36681 , G36689 , G36732 , G36751 , G36761 , G37108 , G37119 , G37129 , G37158 , G37168 , G37188 , G37196 , G37209 , G37234 , G37239 , G37280 , G37293 , G37357 , G37359 , G37413 , G37422 , G37430 , G37442 , G37456 , G37466 , G37475 , G37490 , G37494 , G37501 , G37508 , G37528 , G37536 , G37587 , G37588 , G37641 , G37646 , G37678 , G38058 , G38060 , G38069 , G38076 , G38082 , G38173 , G38178 , G38190 , G38198 , G38272 , G38279 , G38292 , G38293 , G38385 , G38391 , G38613 , G38629 , G38635 , G38643 , G38652 , G38657 , G38666 , G38711 , G38712 , G39465 , G39467 , G39518 , G39533 , G39551 , G39588 , G39589 , G39639 , G39650 , G39657 , G39782 , G39791 , G39800 , G39801 , G39808 , G39809 , G39818 , G39822 , G40016 , G40024 , G40025 , G40116 , G40120 , G40124 , G40126 , G40185 , G40186 , G40190 , G40196 , G40199 , G40203 , G40211 , G40216 , G40227 , G40590 , G40989 , G41058 , G41066 , G41083 , G41098 , G41100 , G41104 , G41108 , G41109 , G41114 , G41119 , G41143 , G41163 , G41192 , G41203 , G41213 , G41218 , G41222 , G41234 , G41244 , G41245 , G41256 , G41267 , G60058 , G60061 , G60062 , G60063 , G60064 , G60065 , G60066 , G60067 , G60069 , G60070 , G60166 , G60169 , G60400 , G60406 , G60415 , G60418 , G60429 , G60430 , G60431 , G60432 , G60435 , G60461 , G60518 , G60520 , G60576 , G60577 , G60582 , G60597 , G60606 , G60631 , G60635 , G60652 , G61307 , G61311 , G61333 , G61337 , G61339 , G61356 , G61367 , G61368 , G61393 , G61394 , G61698 , G61702 , G61706 , G61707 , G61711 , G61715 , G61716 , G61720 , G61725 , G61726 , G61786 , G61795 , G61804 , G61812 , G61814 , G61818 , G61867 , G61872 , G61873 , G61880 , G62011 , G62012 , G62015 , G62020 , G62029 , G62030 , G62037 , G62053 , G62059 , G62062 , G62066 , G62500 , G62502 , G62511 , G62558 , G62564 , G62565 , G62566 , G62626 , G62660 , G62662 , G62682 , G62684 , G62687 , G62747 , G62750 , G62762 , G62804 , G62814 , G62821 , G62828 , G62860 , G62863 , G62867 , G62883 , G62891 , G62892 , G62900 , G62938 , G62943 , G63212 , G63318 , G63328 , G63333 , G63495 , G64500 , G64510 , G64512 , G64789 , G64807 , G64813 , G64854 , G64855 , G64883 , G64891 , G64941 , G65043 , G65172 , G65266 , G65269 , G65275 , G65281 , G65281 , G65289 , G65290 , G65393 , G65395 , G65588 , G65691 , G65692 , G66862 , G66872 , G67257
510(k) numbers
["K101517"]
PMA numbers
not on file
Product code
IZL
Firm FEI
3002495042
Firm city
Algete
Firm state
not on file
Firm country
Spain

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register