Beckman Coulter Inc.: SYNCHRON Systems Modular Creatinine (CREm) Reagent Kit, Part Number: 472525 Systems AQUA CAL 1 and 2, is intended for the quantitative determination of creatinine concentration in human serum, plasma or urine. CREm reagent, when used in conjunction with SYNCHRON LX System(s), UniCel DxC 800 System and SYNCHRON Systems AQUA CAL 1 and 2, is intended for the quantitative determination of creatinine concentration in human serum, plasma or urine.

Recall number
Z-2389-2010
Recalling firm
Beckman Coulter Inc.
Product
SYNCHRON Systems Modular Creatinine (CREm) Reagent Kit, Part Number: 472525 Systems AQUA CAL 1 and 2, is intended for the quantitative determination of creatinine concentration in human serum, plasma or urine. CREm reagent, when used in conjunction with SYNCHRON LX System(s), UniCel DxC 800 System and SYNCHRON Systems AQUA CAL 1 and 2, is intended for the quantitative determination of creatinine concentration in human serum, plasma or urine.
Status
Terminated
Initiated
2010-08-02
Posted
2010-09-09
Terminated
2012-07-02
Reason
The recall was initiated after Beckman Coulter confirmed reports that black / brown particulates were observed in some bottles of the Creatinine Alkaline Buffer (Buffer lot M911427 and M911518) in modular Creatinine (CREm) reagent kit lots M911428 and M911519. Beckman Coulter have had reports of erroneous results when using the lots of Creatinine Alkaline Buffer listed above. Single creatinine d
Root cause
Pending
Action
The firm, Beckman Coulter, sent a "URGENT: PRODUCT CORRECTIVE ACTION" letter dated August 3, 2010, to all customers. The letter describes the product, problem, and action to be taken by customer. The customers were instructed to: (1) Please discontinue use of Creatinine (CREm) reagent kit lots M911428 and M911519 and discard appropriately. (2) Please contact Beckman Coulter at (800) 526-3821 in the United States or (800) 463-7828 in Canada to discuss CREm kit M911428 and M911519 replacement options. Outside the United States and Canada, please contact your local Beckman Coulter Representative. (3) Please complete and return the enclosed response form within 10 days so that we may maintain our records. (4) Please share this information with their laboratory staff and retain this notification as part of their Quality System documentation. If you have forwarded the affected products listed above to another laboratory, please provide a copy of this letter to them. If customers have any questions regarding this Product Corrective Action letter, they were instructed to please contact Beckman Coulter's Customer Support Center (Call Center) on the Beckman Coulter website at www.BeckmanCoulter.com/Customer Support/Support Requests, or call (800) 854-3633 in the United States and Canada.
Quantity
3106 units
Distribution
Worldwide Distribution - USA including the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KY, KS, LA, MA, ME, MD, MI, MN, MO, MS, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, VA, WA, WI, WV, and WY and country of Canada
Lot, serial or model codes
Lot Number(s): M911428 & M911519
510(k) numbers
["K965240"]
PMA numbers
not on file
Product code
CGX
Firm FEI
2050012
Firm city
Brea
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register