GE Medical Systems, SCS: Innova IGS 630/520/530/540 Interventional Fluoroscopic X-ray Systems
- Recall number
- Z-2387-2018
- Recalling firm
- GE Medical Systems, SCS
- Product
- Innova IGS 630/520/530/540 Interventional Fluoroscopic X-ray Systems
- Status
- Terminated
- Initiated
- 2018-03-29
- Posted
- 2018-10-05
- Terminated
- 2021-06-15
- Reason
- The Discovery IGS and Innova IGS systems may experience X-ray abort errors which could result in the loss of imaging mode before or during fluoroscopic use.
- Root cause
- Process control
- Action
- GE Healthcare notified customers on approximately 03/29/2018, via an URGENT MEDICAL DEVICE CORRECTION letter. Safety instructions included that before each use, to ensure the IGS system is fully functional as indicated in the GE product labeling and follow each facility's established procedures before and during each use to manage the patient if a repeated X-ray abort error occurs during the procedure or in the case a complete loss of fluoroscopic viewing capability may occur. If the problem persists, customers are instructed to contact their local GE Healthcare Service Representative. Additionally, customers are instructed that GE Healthcare will correct all affected products at no cost and a GE Healthcare representative will contact each customer to arrange for the correction. Questions or concerns regarding the notification can be directed to GE Healthcare Service at 1-800-437-1171 or the local Service Representative.
- Quantity
- 102 devices total
- Distribution
- US distribution: AR, CA, CA, FL, GA, IA, IL, NC, NH, NJ, NM, NY, PA, SC, TN, A,WI, WV. International Distribution: Argentina, Brazil, China, Colombia, Egypt, France, India, Indonesia, Iran, Italy, Japan, Kuwait, Lithuania, Myanmar, Palestine, Poland, Russia, Saudi Arabia, Singapore, Spain, Tunisia, Turkey, United Kingdom, Vietnam.
- Lot, serial or model codes
- a. Innova IGS 630, Model Number 5730860 - Lot Number (System ID): 00000B3-18-001 (401BIPLANE1), 00000B3-17-032 (505841CATH3), 00000B3-17-029 (956632IGS630), 00000B3-17-030 (YV4502), 00000B3-18-002 (941629IGS630), 00000B3-17-031 (082416160071), 00000B3-18-003 (GON 4601349); b. Innova IGS 520, Model Number 5731194 - Lot Number (System ID): 00000M2-18-008 (410337CATH3), 00000M2-17-090 (603663DIGS), 00000M2-17-088 (603433IGS520), 00000M2-18-004 (423493CATH1), 00000M2-18-002 (414649IGS5), 00000M2-18-003 (414649IGS6), 00000M2-17-087 (M2505310), 00000M2-18-009 (00197VAS02), 00000M2-17-091 (082416040075), 00000M2-18-012 (GON 4607591), 00000M2-18-007 (GON 4607578), 00000M2-18-010 (GON 4607585), 00000M2-18-018 (GON 4592554), 00000M2-17-089 (083016866794118), 00000M2-18-001 (083016865380618), 00000M2-18-005 (GON 4608047), 00000M2-18-019 (GON 4596530), 00000M2-18-013 (GON 4610633), 00000M2-17-092 (A5187111), 00000M2-18-015 (FEG51905), 00000M2-18-006 (GON 4588912); c. Innova IGS 530, Model Number 5731195 - Lot Number (System ID): 00000M3-17-146 (479441CV4), 00000M3-18-007 (352726CATH1), 00000M3-17-136 (815759MCHCAR1), 00000M3-17-137 (856342INIGS), 00000M3-17-132 (856342IGS), 00000M3-17-148 (724658IGS530), 00000M3-18-004 (864573CL1), 00000M3-17-144 (901516INIGS), 00000M3-18-016 (GON 4480942), 00000M3-17-143 (304256INGS), 00000M3-18-005 (XV7945AR), 00000M3-18-005 (GON 1589001), 00000M3-18-011 (GON 1599555), 00000M3-17-150 (XVM317150), 00000M3-17-150 (4555829XV11), 00000M3-18-009 (GON 4603874), 00000M3-18-006 (XV709531BU8), 00000M3-18-003 (GON 4587457), 00000M3-18-003 (XVM3-18-003), 00000M3-18-023 (082416110059), 00000M3-17-135 (082416160068), 00000M3-18-018 (082416120077), 00000M3-17-142 (082416120092), 00000M3-17-152 (082416030051), 00000M3-18-024 (GON 4619686), 00000M3-18-022 (GON 4615780), 00000M3-17-140 (082416160075), 00000M3-17-139 (082416100140), 00000M3-18-021 (GON 4611356), 00000M3-17-141 (082416090100), 00000M3-17-145 (082416100152), 00000M3-18-020 (082416140030), 00000M3-18-017 (082416230041), 00000M3-17-134 (082416250025), 00000M3-17-151 (082416250024), 00000M3-17-133 (082416270031), 00000M3-18-008 (GON 4537575), 00000M3-17-147 (083016246178617), 00000M3-17-154 (5973XR0127), 00000M3-17-153 (5512XR0234), 00000M3-18-013 (GON 4601349), 00000M3-18-015 (GON 4604677), 00000M3-18-002 (6073XR0067), 00000M3-18-012 (GON 4496858), 00000M3-18-001 (RU1079VA01), 00000M3-17-149 (16055VAS01), 00000M3-17-138 (5973XR0121), 00000M3-17-155 (5973XR0112); d. Innova IGS 540 Model Number 5731196- Lot Number (System ID): 00000M4-17-053 (0770385IGS540), 00000M4-18-001 (716631INNOVAIGS), 00000M4-17-052 (215481IGS1), 00000M4-18-002 (215949IGS1), 00000M4-18-004 (843792IGS540), 00000M4-17-048 (253874IR), 00000M4-18-003 (GON 4537782), 00000M4-17-049 (082416040090), 00000M4-18-005 (GON 4619684), 00000M4-17-047 (082416240024), 00000M4-17-050 (082416100144), 00000M4-17-051 (PL1107VA05)
- 510(k) numbers
- ["K111209","K122457"]
- PMA numbers
- not on file
- Product code
- OWB
- Firm FEI
- 1000284313
- Firm city
- Buc
- Firm state
- not on file
- Firm country
- France
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28