Skip to the content

FDA Medical Device Recalls (openFDA)

Tedan Surgical Innovations Llc: Sterile Distraction Screws of 3 different sizes: 12mm, 14mm and 16mm sold under TSI, 3COR, DB Surgical, and American Surgical Products brands and Sterile XL Screws of 7 different sizes: 90mm, 100mm, 110mm, 120mm, 130mm, 140mm, 150mm sold under TSI brand Sterile Distraction Screws are intended for the distraction in the anterior approach to the cervical spine. Sterile XL Screws are intended for the fixation of the blade to the vertebral body to prohibit drifting in the extreme lateral approach to the lumbar spine.

Recall number
Z-2385-2012
Recalling firm
Tedan Surgical Innovations Llc
Product
Sterile Distraction Screws of 3 different sizes: 12mm, 14mm and 16mm sold under TSI, 3COR, DB Surgical, and American Surgical Products brands and Sterile XL Screws of 7 different sizes: 90mm, 100mm, 110mm, 120mm, 130mm, 140mm, 150mm sold under TSI brand Sterile Distraction Screws are intended for the distraction in the anterior approach to the cervical spine. Sterile XL Screws are intended for the fixation of the blade to the vertebral body to prohibit drifting in the extreme lateral approach to the lumbar spine.
Status
Terminated
Initiated
2012-06-11
Posted
2012-09-14
Terminated
2013-07-31
Reason
This letter is to notify you that several of our sterile screw products are being voluntarily recalled. Product sterility may be compromised due to a mechanical failure of the protective packaging tube. Tube end plugs may dislodge creating the potential for the screw to come out of the protective tube and then the sharp end of the screw pierce the outer sterile barrier pouch.
Root cause
Packaging
Action
The firm, TeDan Surgical Innovations (TSI), sent an "URGENT: Medical Device Recall" letter dated June 11, 2012 to their consignees/customers. The letter described the product, problem and actions to be taken. The customers were instructed to immediately examine their inventory and quarantine all products subject to recall and return product to TSI. In addition, if they have further distributed this product, identify all customers and notify them of this product recall at once in writing; and complete and return the enclosed Mandatory Response Form as soon as possible and no later than June 25, 2012 via mail: 12675 W. Airport Blvd., Suite 200, Sugar Land, TX 77478; fax: (713) 726-0846 or email: thernandez@tedansurgical.com. If you have any questions, contact the Business Manager at 713-726-0886, Monday thru Friday, from 8am - 5pm or via email at THERNANDEZ@TEDANSURGICAL.COM.
Quantity
12,065 Sterile Distraction Screws (2413 boxes); 384 Sterile XL Screws (128 boxes) in distribution
Distribution
Worldwide distribution: USA (nationwide) and countries of : Australia and Taiwan.
Lot, serial or model codes
Lot numbers: 019715, 019716, 019942, 019943, 019977, 019978, 020102, 020103, 020104, 020105, 020106, 020107, 020108, 020609, 020782,  021539, 021663, 022097, 022098, 022099, 022122, 022123, 022124, 022371, 022372, 022790, 022791, 022792, 022794, 022795, 022796, 022827, 022828, and 023266.
510(k) numbers
not on file
PMA numbers
not on file
Product code
LXH
Firm FEI
3005726885
Firm city
Sugar Land
Firm state
TX
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register