Skip to the content

FDA Medical Device Recalls (openFDA)

Philips Electronics North America Corporation: Philips DigitalDiagnost Release 4.0.3, Release 4.1/4.1.1. excluding systems with SP1 Product Usage: The DigitalDiagnost is intended for use in generating radiographic images of human anatomy by qualified/trained doctors or technicians.

Recall number
Z-2383-2015
Recalling firm
Philips Electronics North America Corporation
Product
Philips DigitalDiagnost Release 4.0.3, Release 4.1/4.1.1. excluding systems with SP1 Product Usage: The DigitalDiagnost is intended for use in generating radiographic images of human anatomy by qualified/trained doctors or technicians.
Status
Terminated
Initiated
2015-03-23
Posted
not on file
Terminated
2017-01-17
Reason
The system is designed to emit a beep upon termination of an exposure. However, if the system has been powered on for over 12 hours, the system will no longer emit this signal due to a defect in the Microsoft Windows 7 operating system.
Root cause
Radiation Control for Health and Safety Act
Action
Philips Healthcare sent an Urgent Field Safety Notice letter dated June 25, 2015 to customers. The letter identified the affected product, problem and actions to be taken. Philips plans to install a software upgrade in affected systems, which will eliminate this issue. Philips Healthcare Field Service Engineers will schedule an appointment with customers to install the software update. For questions contact your local philips representative.
Quantity
62
Distribution
US Nationwide Distribution to the states of : Arizona, Minnesota, State of Washington, Wyoming and Ohio. Internationally to Canada.
Lot, serial or model codes
DigitalDiagnost
510(k) numbers
["K141736"]
PMA numbers
not on file
Product code
MQB
Firm FEI
1218950
Firm city
Andover
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register