Skip to the content

FDA Medical Device Recalls (openFDA)

Gyrus Acmi, Incorporated: Gyrus ACMI, REF 25 BX, PTFE Coated 3 cm Flexible Tip Guidewire, .035" x 150 cm, Sterile EO, Rx Only, Gyrus ACMI, Inc. 136 Turnpike Road,Southborough, MA 01772-2104 USA Ureteral guidewires are used to establish access in the urinary system for the passage of urological devices. An introducer is first used to establish entry to various portions of the urinary tract (urethra, bladder, ureter), then a guidewire is threaded through the introducer. Following placement of the guidewire, various urological devices can be introduced into the targeted portion of the urinary tract for various therapeutic and diagnostic procedures.

Recall number
Z-2383-2011
Recalling firm
Gyrus Acmi, Incorporated
Product
Gyrus ACMI, REF 25 BX, PTFE Coated 3 cm Flexible Tip Guidewire, .035" x 150 cm, Sterile EO, Rx Only, Gyrus ACMI, Inc. 136 Turnpike Road,Southborough, MA 01772-2104 USA Ureteral guidewires are used to establish access in the urinary system for the passage of urological devices. An introducer is first used to establish entry to various portions of the urinary tract (urethra, bladder, ureter), then a guidewire is threaded through the introducer. Following placement of the guidewire, various urological devices can be introduced into the targeted portion of the urinary tract for various therapeutic and diagnostic procedures.
Status
Terminated
Initiated
2011-04-15
Posted
2011-05-26
Terminated
2012-08-23
Reason
Gyrus ACMI, Inc. wants to bring a recall issue to their customers attention concerning one lot of their 25BX-PTFE Coated 3 cm Flexible Tip Guidewire. As a result of a manufacturing error in the production this lot of guidewires, some units of product were loaded in the dispenser backwards. In such cases, the stiff end of the guidewire would be dispensed from the product first and, if undetected, could be inserted into the patient. Depending on the specific clinical circumstances of the patient, this could potentially cause perforation of the ureter and surrounding anatomy. Therefore Gyrus ACMI has chosen, out of an abundance of caution, to conduct a voluntary recall of this lot of product.
Root cause
Process control
Action
The firm, Olympus ACMI, sent an "Urgent: Medical Device Recall" letter dated April 15, 2011 to its customers. The letter described the product, problem and actions to be taken. The customers were instructed to do the following: 1) Inspect their stock to confirm if any of the affected lot of products remains in their possession 2)Immediately cease any further use of the affected product, remove, quarantine and return it. 3) Call their Gyrus ACMI customer service representative at 1-888-542-7266 to obtain a Returned Goods Authorization 4) Return the product to: Attn: PTFE Guidewire Recall RGA#_________Gyrus ACMI LP, 6845 Wedgwood Court, Maple Grove, MN 55311 5) Complete and return the RECALL REPLY FORM via fax, even if they are returning no product, to: Regulator Affairs (508) 804-2624. The letter states that the customer will receive repeat mailings until they have faxed in the form. Additionally, If they have further distributed the product they are requested to identify and notify their customers at once. If you have any questions, please do not hesitate to contact customer service at 1-888-524-7266.
Quantity
145
Distribution
Nationwide distribution: USA including states of: CA, GA, HI, IN, KY, MT, NY, NC, PA, TX, UT, and WA.
Lot, serial or model codes
Lot Number 252410KC
510(k) numbers
not on file
PMA numbers
not on file
Product code
KNY
Firm FEI
3003790304
Firm city
Southborough
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register