Skip to the content

FDA Medical Device Recalls (openFDA)

3M Company / Medical Division: 3M Red Dot Monitoring Electrode, Catalog #2560 & Novaplus Universal Monitoring Electrode #V2560, used for ECG monitoring.

Recall number
Z-2383-2010
Recalling firm
3M Company / Medical Division
Product
3M Red Dot Monitoring Electrode, Catalog #2560 & Novaplus Universal Monitoring Electrode #V2560, used for ECG monitoring.
Status
Terminated
Initiated
2010-07-27
Posted
2010-09-07
Terminated
2012-04-19
Reason
The sensing element of the 3M Red Dot Monitoring Electrode, Catalog # 2560 and Novaplus Universal Monitoring Electrode #V2560 may have corrosion. In an emergency situation where defibrillation is necessary and electrodes are already in use, the electrodes may malfunction and provide a delayed or no trace response.
Root cause
Process control
Action
3M sent an "Urgent Medical Device Recall" letters and a "Distributor Product Recall Form" dated July 29, 2010 to customers and distributors of the 3M and Novaplus electrodes. The letters described the 3M Red Dot and Novaplus Universal products, problems and actions to be taken by customers. Customers were instructed to examine their inventory and quarantine the affected lots pending return to the firm. After completing and returning the enclosed Customer Product Recall Form to 3M, customers will be issued a Return Authorization Number and provided with instructions for return of the affected devices. Distributors were instructed to notify customers to the end user level and have them return affected lots to the distributor for forwarding to 3M, or forward the customer letter to end-users to instruct them on how to contact 3M for recovery of the device. The 3M Customer Helpline can be contacted at: 1-800-228-3957.
Quantity
11,063 cases (20 pouches/case, 50 electrodes/pouch)
Distribution
Nationwide Distribution: States of CA, FL, IL, KS, KY, MI, MO, NE, NJ, NY, OH, PA, RI, SD, TN, VA, and WI.
Lot, serial or model codes
2560 LOT #'S: 2012-06 CX, 2012-06 CY, 2012-06 CZ, 2012-06 DA, 2012-06 DB, 2012-05 CN, 2012-05 CM, 2012-05 CK, 2012-05 CJ, 2012-05 CI, 2012-05 CH,2012-05 CG, 2012-05 CF, 2012-05 CE, 2012-05 CD, and 2012-05 CC. V2560 NOVATION PRIVATE LABEL LOT: 2012-05 CL.
510(k) numbers
["K970796"]
PMA numbers
not on file
Product code
DRX
Firm FEI
2110898
Firm city
Saint Paul
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register