Skip to the content

FDA Medical Device Recalls (openFDA)

Thera Test Laboratories, Inc.: EL-Anti-SARS-CoV-2 IgM Kit, Catalog #104-129, IVD. Submitted as EUA202129

Recall number
Z-2381-2021
Recalling firm
Thera Test Laboratories, Inc.
Product
EL-Anti-SARS-CoV-2 IgM Kit, Catalog #104-129, IVD. Submitted as EUA202129
Status
Terminated
Initiated
2020-07-30
Posted
not on file
Terminated
2022-02-15
Reason
Device did not meet the updated EUA packet requirements for positive percent agreement, thus, not adequately validated. Submitted as EUA202129
Root cause
No Marketing Application
Action
An undated letter was issued to their customers attached to an email dated 7/30/2020 explaining the reason for recall and requesting the customer stop testing with the kit.
Quantity
11 kits
Distribution
US Nationwide distribution in the states of CA, FL, IL, IN, NC, and TX.
Lot, serial or model codes
Kit lot numbers 05204350, exp. 05/01/2021; 05204367, exp. 05/19/2021; and 06204432, exp. 05/19/2021.
510(k) numbers
not on file
PMA numbers
not on file
Product code
QKO
Firm FEI
1421346
Firm city
Lombard
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register