Skip to the content

FDA Medical Device Recalls (openFDA)

Hospira Inc: Hospira Inc. Symbiq One- and Two- Channel Infuser, List Numbers 16026 and 16027 respectively.

Recall number
Z-2381-2010
Recalling firm
Hospira Inc
Product
Hospira Inc. Symbiq One- and Two- Channel Infuser, List Numbers 16026 and 16027 respectively.
Status
Terminated
Initiated
2010-03-26
Posted
2010-09-08
Terminated
2012-03-30
Reason
If the administrator set is removed prior to the cassette carriage fully opening without closing the slide/roller clamp, unrestricted flow and delivery may result.
Root cause
Device Design
Action
Hospitra Urgent Device Correction Notification mailed on 03/26/2010 via federal express. Tracking done by delivery tracking. Distribution of product continues and has not stopped. Additional notification was issued in 09/2010. A software correction is planned.
Quantity
28485 units
Distribution
Worldwide distribution: USA, Australia, Canada, Colombia, Costa Rica, Malaysia, Mexico, Puerto Rico, Singapore and the UAE.
Lot, serial or model codes
List number configurations: 16026-04-01/03/04, 51/52, 53/54, 79/80, 81/82, 83/84, 87/88, 89/90; 16027-04-01/03/04, 51/52, 53/54, 79/80, 81/82, 83/84, 87/88, 89/90; 16026-27-89/90, 16026-13-25/26, 53/54; 16027-27-89/90; 16027-13-25/26, 53/54.
510(k) numbers
["K041550"]
PMA numbers
not on file
Product code
FRN
Firm FEI
2921482
Firm city
Morgan Hill
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register