Skip to the content

FDA Medical Device Recalls (openFDA)

Medtronic Neuromodulation: Medtronic INDURA 1P Intrathecal Catheter, model 8709SC. 81.4 cm catheter with length markers and guide wire, 15 T-guage introducer needle, Sutureless pump connector with attached 7.6 cm catheter and connector pin, Transparent Strain-relief sleeves (2), Anchors (4).

Recall number
Z-2380-2008
Recalling firm
Medtronic Neuromodulation
Product
Medtronic INDURA 1P Intrathecal Catheter, model 8709SC. 81.4 cm catheter with length markers and guide wire, 15 T-guage introducer needle, Sutureless pump connector with attached 7.6 cm catheter and connector pin, Transparent Strain-relief sleeves (2), Anchors (4).
Status
Terminated
Initiated
2008-06-26
Posted
2008-09-25
Terminated
2011-06-06
Reason
Disconnection of the Medtronic sutureless connector (SC) catheters from the catheter port on the pump, or occlusion between the sutureless pump connector and the catheter port on the pump.
Root cause
Component design/selection
Action
Medtronic sent A Safety Alert dated June 2008, on June 27, 2008, to healthcare professionals, which described the issues of occlusion and disconnect. The letter identifies the affected catheter and revision kit model numbers and the associated implantable infusion pumps. The issues do not involve Medtronic MiniMed insulin pump. The letter refers the HCP to recommendations for Implant Techniques and Recommendations for Patency Verification that are included with the letter. The letter also includes recommendations for managing patients with implanted SC catheters.
Quantity
21,674 total
Distribution
Worldwide Distribution: USA, Puerto Rico, Algeria, Aruba, Austria, Australia, Belgium, Canada, Croatia, Cyprus, Czech Republic, Denmark, Finland, France, French Polynesia, Germany, Greece, Hong Kong, Hungary, Iceland, India, Ireland, Israel, Italy, Jordan, Kuwait, Lebanon, Luxembourg, Malta, Netherlands, Netherlands Antilles, Norway, Poland, Portugal, Reunion, Romania, Russian Federation, San Marino, Saudi Arabia, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Turkey, United Arab Emirates, and United Kingdom.
Lot, serial or model codes
All SC Catheters
510(k) numbers
not on file
PMA numbers
["P860004S081"]
Product code
LKK
Firm FEI
2182207
Firm city
Minneapolis
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.