FDA Medical Device Recalls (openFDA)

Baxter Healthcare Corporation: Baxter prismaflex, HF 1000 PAES membrane, REF 107140 Used with Prismaflex or PrisMax control units for continuous fluid management and renal replacement therapies.

Recall number
Z-2379-2021
Recalling firm
Baxter Healthcare Corporation
Product
Baxter prismaflex, HF 1000 PAES membrane, REF 107140 Used with Prismaflex or PrisMax control units for continuous fluid management and renal replacement therapies.
Status
Terminated
Initiated
2021-07-22
Posted
not on file
Terminated
2024-01-17
Reason
Products labeled as sterile were distributed, but may not have been sterilized. Baxter confirmed Steril Milano observed nonconformances that were not reported to Baxter Data was manipulated to appear to be conforming
Root cause
Vendor change control
Action
The firm notified its consignees of the recall by letter on 07/22/2021. The letter explained the issue and requested the return of the affected product. Distributors were directed to notified their consignees.
Quantity
1976 units
Distribution
US Nationwide distribution.
Lot, serial or model codes
Lot Numbers: 20B2330M
510(k) numbers
["K042938"]
PMA numbers
not on file
Product code
KDI
Firm FEI
1417572
Firm city
Deerfield
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register