Stryker Howmedica Osteonics Corp.: Stryker Orthopaedics Modular Handle Triathlon Instruments, Non-Sterile, REF 6541-4-808, Lot 123, NDC M546654148080S Product Usage: Is a reusable instrument that is part of the Triathlon Primary total knee arthroplasty (TKA) system as well as the Triathlon Partial Knee Replacement (PKR) system. The use of the handles in Primary TKA is optional and may be used to: (1) Assist in stabilizing the Triathlon Femoral Sizer, (2) Aid in stabilization during use and removal of the Triathlon 4:1 Cutting Block and (3) Aid in assembly of the Anterior MIS Distal Resection Guide to the Anterior Skim Cut Guide. Per the Triathlon PKR surgical protocol, the handle is required for a PKR surgery and is used to: (1) Assemble and disassemble the Femoral Trial Drill Guide into the Femoral Trial and (2) Remove the Peg Trial from the Femoral Trial.

Recall number
Z-2377-2016
Recalling firm
Stryker Howmedica Osteonics Corp.
Product
Stryker Orthopaedics Modular Handle Triathlon Instruments, Non-Sterile, REF 6541-4-808, Lot 123, NDC M546654148080S Product Usage: Is a reusable instrument that is part of the Triathlon Primary total knee arthroplasty (TKA) system as well as the Triathlon Partial Knee Replacement (PKR) system. The use of the handles in Primary TKA is optional and may be used to: (1) Assist in stabilizing the Triathlon Femoral Sizer, (2) Aid in stabilization during use and removal of the Triathlon 4:1 Cutting Block and (3) Aid in assembly of the Anterior MIS Distal Resection Guide to the Anterior Skim Cut Guide. Per the Triathlon PKR surgical protocol, the handle is required for a PKR surgery and is used to: (1) Assemble and disassemble the Femoral Trial Drill Guide into the Femoral Trial and (2) Remove the Peg Trial from the Femoral Trial.
Status
Terminated
Initiated
2016-06-23
Posted
not on file
Terminated
2019-03-27
Reason
Stryker Orthopaedic is voluntarily recalling the Triathlon Modular Handle because analysis revealed that the press fit specifications between the dowel pin and the mating hole in the shaft were not being achieved potentially resulting in disassociation from the instrument.
Root cause
Device Design
Action
Stryker sent an Urgent Product Recall letter dated June 24, 2016 to Branches/Agencies by e-mail and Business Reply Forms was sent to Branches/Agents/Hospital Risk Management via UPS (with return receipt) on June 24, 2016. The letter identified the affected product, problem and actions to be taken. For questions call (201) 831-6693.
Quantity
12,469 units
Distribution
Worldwide Distribution - US Nationwide
Lot, serial or model codes
Catalog No. 6541-4-808 Lot Nos.: SB1K14, SB1K14A, SB1K14A1, SB1K14P, SB1K15, SB1K15A, SB1K15A1, SB1K15A2, SB1M25, SB1M25A, SB1M25A1, SB1M25A2, SB1M25A2X1, SB1N02, SB1N02A, SB1N02A1, SB1N02A2, SB1N02A3, SB1N02X1, SB1T05, SB1T05A, SB1T05A1, SB1T05AX1, SB1T05AX2, SB1T05KSB1V01, SB1V01K, SB1V01L, SB1V01X1, SB1V01X2, SB1V01X3, SB1W14, SB1W14K, SB1W14X1, SB2T19, SB2T19A, SB2T19D, SB2T19DX1, SB2T19DX2, SB2T19J, SB2T19K, SB2T19KX1, SB2T19KX2, SBT19X3, SB2V87, SB2V87A, SB2V87J, SB2V87JX1, SB2V87K, SB2V87X, SB2V87X1, SB2W23, SB2W23G, SB2W23X, SB2W23X1, SB3A20, SB3A20A, SB3A21, SB3A21D, SB3A21M, SB3A21X, SB3A22, SB3A22A1, SB3A22X1, SB3A55, SB3A55A, SB3A55A2, SB3A55AX1, SB3A55M, SB3A55X1, SB3H25, SB3H25D, SB3H25G, SB3H25X1, SB3H26, SB3H26D, SB3H26DX1, SB3H26X2, SB3K09, SB3K09X1, SB3K10, SB3K10L, SB3K10M, SB3K10MX1, SB3K10X1, SB3L19, SB3L19D, SB3L19K, SB3L19L, SB3L19P, SB3L19P1, SB3L20, SB3L20K, SB3L45, SB3L45A, SB3L45A1, SB3L45A2, SB3L45D, SB3L45DX1, SB3L45L, SB3L45T, SB3M17, SB3M17A, SB3M17K, SB3M17M, SB3M18, SB3M18A, SB3M18AX1, SB3M18D, SB3M18M, SB3M18R, SB3M19, SB3M19A, SB3M19A1, SB3M19S, SB3M19T, SB3M19W, SB3N46, SB3N46D, SB3N46P, SB3N46X1, SB4C12, SB4C12X1, SB4C37, SB4C37X1, SB4C38, SB4C38A, SB4C38D, SB4E18, SB4E181, SB4E181X1, SB4E18A, SB4E18D, SB4E18T, SB4H79, SB4H79A, SB4H79M, SB4L05, SB4L05A, SB4N17, SB4N17E, SB4N17M, SB4N17M1, SB4N18, SB4N18A, SB4N18A1, SB4N18D, SB4N18M, SB4N18T, SB4S01, SB4S01D, SB4S01J, SB4V07, SB4V07L, SB4V07M, SB4V07MM, SB5A02, SB5A02A, SB5A02A1, SB5A55
510(k) numbers
["K143393"]
PMA numbers
not on file
Product code
JWH
Firm FEI
3003070421
Firm city
Mahwah
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register