Skip to the content

FDA Medical Device Recalls (openFDA)

Covidien, LP: Palindrome Precision RT Chronic Catheter Kit Symmetrical Tip, Reverse-Tunneled 15 Fr/Ch (5.0 mm) x 55 cm, Item Code 8888541055P

Recall number
Z-2375-2021
Recalling firm
Covidien, LP
Product
Palindrome Precision RT Chronic Catheter Kit Symmetrical Tip, Reverse-Tunneled 15 Fr/Ch (5.0 mm) x 55 cm, Item Code 8888541055P
Status
Open, Classified
Initiated
2021-05-28
Posted
not on file
Terminated
not on file
Reason
This voluntary withdrawal is being conducted to implement point-of-use labeling on each device and enhancements to the Instructions for Use (IFU).
Root cause
Labeling design
Action
A Medical Device Market Withdrawal notification letter dated 5/28/21 was sent to customers. Actions being taken by Medtronic 1. Medtronic is developing point-of-use labeling reinforcing the important user assembly steps contained within the IFU related to the hub/back end assembly procedure. 2. The IFU for Palindrome" RT and Palindrome" Precision RT chronic hemodialysis catheters will be updated to enhance the detailed instructions for users related to the steps required for hub/back end assembly. Actions you should take 1. Please immediately quarantine and discontinue use of affected item codes and lots listed in Attachment A. 2. Please return affected product as indicated below. All unused products from the affected item codes and lots must be returned. 3. If you have distributed the Palindrome" RT and Palindrome" Precision RT chronic hemodialysis catheters listed in Attachment A, please promptly forward the information from this letter to those recipients. 4. Complete the Product Return Form even if you do not have inventory. Purchased directly from Medtronic 1. Complete the Product Return Form and fax or email to Medtronic. 2. Ship affected product with RGA number provided by Customer Service to: Medtronic Attn: Field Returns 195 McDermott Road North Haven, CT 06473 USA Fax to 800-895-6140 or email to rs.covidienfeedbackcustomerservice@medtronic.com Credit for returned affected product will be issued based on the RGA number. Exception: Customers with zero inventory, fax to 651-367-2624 or email to rs.gmbfcamitg@medtronic.com Purchased from a distributor 1. Complete the Product Return Form and fax or email to Medtronic 2. Contact your distributor directly to arrange for return of product back to your distributor. Fax to 651-367-2624 or email to rs.gmbfcamitg@medtronic.com This action is being taken with the knowledge of the FDA and other regulatory authorities. We request that you contact Medtronic if you experienced a quality problem or adverse event. Ema
Quantity
not on file
Distribution
Worldwide distribution.
Lot, serial or model codes
Lot Numbers: 1726300131 1805900130 1911900083 2019900069 1731900198 1813600105 1926100249 2023300142 1800300137 1827600140 1926100293 2023300118 1802400067 1835100109 2002100137 2034400093
510(k) numbers
["K123196"]
PMA numbers
not on file
Product code
MSD
Firm FEI
1219930
Firm city
North Haven
Firm state
CT
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register