Skip to the content

FDA Medical Device Recalls (openFDA)

Teleflex Medical: Servo Humidifier 163, Model 01-06-8125-8, REF 6419365 Product Usage: A heat and moisture condenser (artifical nose) is a device intended to be positioned over a tracheotomy (a surgically created opening in the throat) or tracheal tube (a tube inserted into the trachea) to warm and humidity gases breathed in by the patient.

Recall number
Z-2374-2016
Recalling firm
Teleflex Medical
Product
Servo Humidifier 163, Model 01-06-8125-8, REF 6419365 Product Usage: A heat and moisture condenser (artifical nose) is a device intended to be positioned over a tracheotomy (a surgically created opening in the throat) or tracheal tube (a tube inserted into the trachea) to warm and humidity gases breathed in by the patient.
Status
Terminated
Initiated
2015-12-03
Posted
2016-08-09
Terminated
2017-04-14
Reason
Cracks may occur at the 15mm and 22mm ISO female connector.
Root cause
Release of Material/Component prior to receiving test results
Action
Inmed, a division of Teleflex sent an Urgent Field Safety Notice letter dated December 3, 2015 to all affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to check inventory, cease use and distribution of stock of the affected product batch and quarantine immediately.. The letter included an acknowledgment form which was to be completed and returned your MAQUET representative. For further information or support concerning this issue, contact your local MAQUET representative at fieldactions@maquet.com. For questions contact your local sales representative or Customer Service.
Quantity
2,700 ea.
Distribution
US Nationwide Distribution in the states of FL, KS and WV.
Lot, serial or model codes
Lots 201450, 201451, 201452, 201504
510(k) numbers
not on file
PMA numbers
not on file
Product code
BYD
Firm FEI
3005747797
Firm city
Morrisville
Firm state
NC
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register