Covidien LP: MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 13 cm, Curved Extensions, PASS Tray REF: 8888103001HP

Recall number
Z-2373-2023
Recalling firm
Covidien LP
Product
MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 13 cm, Curved Extensions, PASS Tray REF: 8888103001HP
Status
Open, Classified
Initiated
2023-06-28
Posted
2023-08-11
Terminated
not on file
Reason
Catheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or excessive MDX may dislodge from the catheter lead to full catheter obstruction resulting in delay to treatment and partial obstruction resulting in reduced flow, or particulate dislodgement that may result in delay to treatment, hemolysis, embolism/embolus or thrombosis/thrombus
Root cause
Process control
Action
Medtronic issued Urgent Medical Device Recall letter beginning on June 28, 2023. Letter states reason for recall, health risk and action to take: To help you identify if you have affected product, please visit our website www.Medtronic.com/Mahurkar-Triple-Lumen-Catheter-Recall. Here you will find a tool to help you determine if the product you have is affected by this recall. 2. Immediately quarantine and discontinue use of all unused Mahurkar TM Acute Triple Lumen Catheters and Mahurkar TM Acute High Pressure Triple Lumen Catheters referenced in Attachment B - List of affected Lot numbers (see Attachment A for guidance to identify impacted product). Note: This recall does not include Mahurkar Elite Catheters. 3. Complete the online Customer Confirmation Form using the link below within 3 business days of receiving this notice. Replying promptly will confirm your receipt of the notification and prevent you from receiving repeat notices. Link: https://www.novasyte.com/medtronic/fa1333 Note: Please complete the online Customer Confirmation Form even if you DO NOT have affected product. 4. If you have affected product, it must be returned for processing. Upon completion of the online Customer Confirmation Form, Customer Service will contact you with an RGA number to return all unused affected Mahurkar TM Acute Triple Lumen Catheters and Mahurkar TM Acute High Pressure Triple Lumen Catheters from your inventory to Medtronic. 5. This notice should be passed on to all those who need to be aware within your organization or to any organization including but not limited to Nephrologists, Intensivists, physicians, renal nurses, critical care nurses, or other dialysis staff where the potentially affected devices have been transferred. Please maintain a copy of this notice in your records. Medtronic has partnered with Novasyte, an IQVIA company, to assist in this recall. For assistance regarding online response processing or product return for this recall, please contact Novas
Quantity
299 units
Distribution
US Nationwide - Worldwide Distribution Foreign: Argentina Austria Bahrain Bangladesh Belgium Brazil Brunei Darussalam Canada Cayman Islands China Costa Rica Czech Republic Denmark Dominican Republic Finland France French Polynesia Germany Greece Hungary India Iran, Islamic Republic Of Ireland Italy Kuwait Malaysia Maldives Mauritius Mexico Nepal Netherlands New Caledonia Nicaragua Norway Philippines Poland Portugal Puerto Rico Saudi Arabia Singapore Slovakia South Africa Spain Sri Lanka Sweden Switzerland Thailand Turkey United Arab Emirates United Kingdom
Lot, serial or model codes
GTIN: 0884521128010 LOT Numbers: 2133700228
510(k) numbers
["K020089","K102605"]
PMA numbers
not on file
Product code
NIE
Firm FEI
1282497
Firm city
Mansfield
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register