Skip to the content

FDA Medical Device Recalls (openFDA)

Customed, Inc: FOLEY CATHETER KIT 1116FR. - CUSTOM ED- (1) WRAPPER 24" X 24" (1) PVP TRIPLE SWABSTICK IODOPHOR (2) GLOVE MED FREETOUCH VYNIL P/F (1) JELLY LUBRICATING 5 GRAM (1) TOWEL 3PLY POLY 17" X 18" (2) SWABSTICK DRY COTTON (1) DRAPE 4" FENESTRATED 18" X 26" (1) CATHETER 16FR. FOLEY, 2 WAY,5cc, Silicone Coated (1) SYRINGE 10cc WATERPREFIL UF (1) BAG URINARY DRAINAGE 2000ML ANTI-REFLUX UF (1) LABEL ID PATIENT (1) TRAY ORGANIZER (1) SPECIMEN CONTAINER 4oz. W/LID & LABEL EO sterilized surgical convenient packs packaged into polyester/polyethylene breather bag with Tyvek double vents.

Recall number
Z-2373-2014
Recalling firm
Customed, Inc
Product
FOLEY CATHETER KIT 1116FR. - CUSTOM ED- (1) WRAPPER 24" X 24" (1) PVP TRIPLE SWABSTICK IODOPHOR (2) GLOVE MED FREETOUCH VYNIL P/F (1) JELLY LUBRICATING 5 GRAM (1) TOWEL 3PLY POLY 17" X 18" (2) SWABSTICK DRY COTTON (1) DRAPE 4" FENESTRATED 18" X 26" (1) CATHETER 16FR. FOLEY, 2 WAY,5cc, Silicone Coated (1) SYRINGE 10cc WATERPREFIL UF (1) BAG URINARY DRAINAGE 2000ML ANTI-REFLUX UF (1) LABEL ID PATIENT (1) TRAY ORGANIZER (1) SPECIMEN CONTAINER 4oz. W/LID & LABEL EO sterilized surgical convenient packs packaged into polyester/polyethylene breather bag with Tyvek double vents.
Status
Terminated
Initiated
2014-05-20
Posted
2014-08-29
Terminated
2017-08-14
Reason
Customed has determined that there is the possibility that packaging integrity may be compromised on different catalog numbers (multiple lots). This could result in an injury to the patient due to product contamination or loss of sterility condition.
Root cause
Package design/selection
Action
Customed Inc. sent a Recall Notification letter dated June 3, 2014 via email and followed by letter to all affected customers. The letter identified the affected products, problem and actions to be taken. The letter informed customers if the affected products are in their inventory to: 1. Remove products from sale (quarantine them) and cease distribution. 2. Complete and return the enclosed acknowledgement form to Customed Inc .Recall Coordinator. Affected products should be returned to Customed Inc for credit. Customers with questions were instructed to contact Customed Inc. Recall Coordinator at 1-787-801-0100 Ext. 7540
Quantity
27 lots; 67410 units (multiple units per lot)
Distribution
Worldwide Distribution - US (nationwide) in the state of Florida including Puerto Rico (USA); & USVI (St John) and the countries of Tortola, Trinidad, Dominican Republic, Costa Rica and Mexico.
Lot, serial or model codes
Product code 900-1127, 27 lots: 111082044 111092578 111102824 111112983 111123318 112020311 112030602 112041178 112072767 112083205 112093744 113015572 113025906 113036449 113047145 113068141 113088830 113099339 131110626 131210624 140111138 140211532 140211883 140412442 140412828 140513066 140513253
510(k) numbers
not on file
PMA numbers
not on file
Product code
NWR
Firm FEI
2648727
Firm city
Fajardo
Firm state
not on file
Firm country
PR

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register