FDA Medical Device Recalls (openFDA)
Elite Biomedical Solutions LLC: Alaris Medley LVP Frame Membrane. Intended use to hold platen assembly and help protect the bezel assembly's internal components from fluid intrusion. This part is used in the Alaris Medley LVP 8100 infusion pump.
- Recall number
- Z-2371-2015
- Recalling firm
- Elite Biomedical Solutions LLC
- Product
- Alaris Medley LVP Frame Membrane. Intended use to hold platen assembly and help protect the bezel assembly's internal components from fluid intrusion. This part is used in the Alaris Medley LVP 8100 infusion pump.
- Status
- Terminated
- Initiated
- 2015-06-03
- Posted
- 2015-08-20
- Terminated
- 2017-03-07
- Reason
- Administration of inappropriate quantities of fluid can result, with the potential to cause injury or death.
- Root cause
- Other
- Action
- Elite Biomedical Solutions, sent a "Product Advisory Notices" letter dated May 21, 2015 to their customers. On June 3, 2015 the firm sent an Urgent Medical Device Recall letter to their customers. The letter identified the affected product , problem and actions to be taken. And on June 12, 2015 a press release was issued via ECRI (Emergency Care Research Institute) to all hospitals in the US. The notification letter requested customers to: 1) inspect and quarantine affected products, 2) identify your customers and notify them at once of this product recall, and 3) complete and return the enclosed response forms. For any questions, call Elite Biomedical Solutions, LLC, Quality Manager, at 1-855-291-6701.
- Quantity
- 609 pieces
- Distribution
- Nationwide Distribution in the states of AZ, CA, GA, IL, IN, KS, KY, MA, MD, MN, MO, MS, NC, NY, OH, PA, SC, TN, TX, VA & WI.
- Lot, serial or model codes
- P/N TC10006587 and 10013801 Lot Numbers 022015502 and 0421151000.
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- FRN
- Firm FEI
- 3009712113
- Firm city
- Cincinnati
- Firm state
- OH
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28