Philips Medical Systems: Philips Telemonitoring System Software (M3811B) used with Philips Telemonitoring System (M3810B) Revision: B.02.07 Prescriptive medical device used to automatically collect and transmit medical information (weight, blood pressure, non-diagnostic ECG) over phone lines between provider and patient.

Recall number
Z-2369-2008
Recalling firm
Philips Medical Systems
Product
Philips Telemonitoring System Software (M3811B) used with Philips Telemonitoring System (M3810B) Revision: B.02.07 Prescriptive medical device used to automatically collect and transmit medical information (weight, blood pressure, non-diagnostic ECG) over phone lines between provider and patient.
Status
Terminated
Initiated
2008-07-22
Posted
2008-09-18
Terminated
2012-05-16
Reason
Software: If Weight Limits edited, values will return to default values that were deleted
Root cause
Software design
Action
Affected customers will receive the Urgent Medical Device Correction notification on 7/22/08 via certified mail. Delivery receipt will be tracked. The Urgent Medical Device Correction notification instructs customers to verify that daily and weekly weight limits are set or not set as intended while they await the correction, which will come in the form of a new version of software and will simultaneously update all systems. Philips has corrected the software defect and is in the process of validating the correction and releasing the corrected software as part of a scheduled release. If further information or support is needed, customers should contact the Telehealth Customer Service Line at 1-866-246-7316.
Quantity
49 units
Distribution
Nationwide Distribution.
Lot, serial or model codes
Software Revision: B.02.07
510(k) numbers
["K041674"]
PMA numbers
not on file
Product code
DQA
Firm FEI
1218950
Firm city
Andover
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register