FDA Medical Device Recalls (openFDA)

Hitachi Medical Systems America Inc: Oasis Coil Extension Cable used with Hitachi Oasis MRI System Product Usage: The Oasis MR system is an imaging device and is intended to provide the physician with physiological and clinical information, obtained non-invasively and without the use of ionizing radiation. The Oasis system produces transverse, coronal, sagittal, oblique, and curved cross-sectional images that display the internal structure of the head, body, or extremities. The images produced by the MR systems reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. The NMR properties that determine the image appearance are proton density, spin-lattice relaxation time (T1), spin-spin relaxation time (T2), and flow. When interpreted by a trained physician, these images provide information that can be useful in diagnosis determination. The Coil Extension Cable that is the subject of the removal action is an accessory that is used in conjunction with certain anatomy coils to aid in patient positioning.

Recall number
Z-2366-2015
Recalling firm
Hitachi Medical Systems America Inc
Product
Oasis Coil Extension Cable used with Hitachi Oasis MRI System Product Usage: The Oasis MR system is an imaging device and is intended to provide the physician with physiological and clinical information, obtained non-invasively and without the use of ionizing radiation. The Oasis system produces transverse, coronal, sagittal, oblique, and curved cross-sectional images that display the internal structure of the head, body, or extremities. The images produced by the MR systems reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. The NMR properties that determine the image appearance are proton density, spin-lattice relaxation time (T1), spin-spin relaxation time (T2), and flow. When interpreted by a trained physician, these images provide information that can be useful in diagnosis determination. The Coil Extension Cable that is the subject of the removal action is an accessory that is used in conjunction with certain anatomy coils to aid in patient positioning.
Status
Terminated
Initiated
2014-04-25
Posted
2015-08-10
Terminated
2018-08-24
Reason
The firm received a complaint stating the Coil Extension Cable accessory would overheat and be a burn hazard if touched.
Root cause
Nonconforming Material/Component
Action
Hitachi sent an Urgent: Device Removal Notice letter date April 24, 2014 to their customers. The letter identified the affected product, problem and actions to be taken. The letter informed customers to discontinue using the Coil Extension Cable and that a Field Service Engineer will schedule a visit to remove the cable. For questions call 800-800-3106 ext 3720.
Quantity
173 units
Distribution
US Nationwide in the states: AL, AZ, CA, CO, CT, DC, DE, FL, GA, IA, ID, IL, IN, KS, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NJ, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI & WY and the country of Mexico.
Lot, serial or model codes
Serial #'s: M001to M108, M110 to M175, M951 & M952.
510(k) numbers
["K072279"]
PMA numbers
not on file
Product code
MOS
Firm FEI
1528028
Firm city
Twinsburg
Firm state
OH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register