Boston Scientific Corporation: Boston Scientific Platinum Plus Guidewire. Material No. [outer carton label] H74917541, Catalog # 1754, Material No. [Inner Pouch Label] H74917540. Sterilized with ethylene oxide gas. The Platinum Plus Guidewire has been designed to provide torsional control. The guidewire has a shapeable and highly radiopaque distal platinum tip. Legal Manufacturer: Boston Scientific Corporation, One Boston Scientific Place, Natick, MA 01760-1537, USA.

Recall number
Z-2365-2008
Recalling firm
Boston Scientific Corporation
Product
Boston Scientific Platinum Plus Guidewire. Material No. [outer carton label] H74917541, Catalog # 1754, Material No. [Inner Pouch Label] H74917540. Sterilized with ethylene oxide gas. The Platinum Plus Guidewire has been designed to provide torsional control. The guidewire has a shapeable and highly radiopaque distal platinum tip. Legal Manufacturer: Boston Scientific Corporation, One Boston Scientific Place, Natick, MA 01760-1537, USA.
Status
Terminated
Initiated
2008-06-19
Posted
2008-09-21
Terminated
2010-10-13
Reason
Boston Scientific is initiating a voluntary recall of the Platinum Plus Coronary Guidewire. Through our internal inspection process, we identified the potential that the PTFE (polytetrafluoroethylene) coating on the Platinum Plus Coronary Guidewires could be damaged in certain locations. This damage of the coating creates the potential for small particles of the PTFE coating to detach from the wi
Root cause
Process control
Action
An urgent Voluntary medical Device Recall or Correction - Immediate Action Required letter was sent to all affected customers. The letter describes the issue and clinical effects. A list of affected devices is attached to the letter. The customer is to cease use or distribution of affected product, and return it immediately to Boston Scientific. Return directions are provided in an attached document. A Reply Verification Tracking Form is to be completed and faxed back to Customer Service. If any product is being returned the completed form is to accompany the returned product. Contact Boston Scientific Corporation at 1-763-494-1700 for assistance.
Quantity
22,397 single units total (18,261 US, 4,136 OUS)
Distribution
AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, ME, MI, MN,MO, MS, MT, NC, ND, NE, NH, NJ, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, SI, WV, and WY. OUS locations to include: Balkans, Finland, Germany, Great Britain, Hong Kong, Ireland, Israel, Italy, Mexico, Netherlands, Norway, Romania, Saudia Arabia, Spain, Sweden, Switzerland, and United Arab Emirates
Lot, serial or model codes
Lot / Batch #: 7873372, 7909303, 7987688, 8025769, 8049696, 8202657, 8203318, 8260511, 8289582, 8351041, 8372192, 8372883, 8410171, 8418162, 8432959, 8506417, 8533771, and 11643756.
510(k) numbers
["K935997"]
PMA numbers
not on file
Product code
DQX
Firm FEI
3002095335
Firm city
Maple Grove
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register