Ortho-Clinical Diagnostics: VITROS Chemistry Product AMYL Slides Reagent, AMYLASE/5 PACK/300 SLDS, Product Code 1202670, UDI # 10758750008513 - Product Usage: For in vitro diagnostic use only to quantitatively measure amylase activity in serum, plasma, and urine using VITROS 250/350/5, 1 FS/4600/XT 3400 Chemistry Systems and VITROS 5600/XT 7600 Integrated Systems.

Recall number
Z-2364-2020
Recalling firm
Ortho-Clinical Diagnostics
Product
VITROS Chemistry Product AMYL Slides Reagent, AMYLASE/5 PACK/300 SLDS, Product Code 1202670, UDI # 10758750008513 - Product Usage: For in vitro diagnostic use only to quantitatively measure amylase activity in serum, plasma, and urine using VITROS 250/350/5, 1 FS/4600/XT 3400 Chemistry Systems and VITROS 5600/XT 7600 Integrated Systems.
Status
Terminated
Initiated
2020-05-11
Posted
not on file
Terminated
2021-09-22
Reason
Multiple coatings of VITROS Chemistry Products AMYL Slides are showing imprecision on control fluids when compared to the expected within lab SD as listed on the Performance Verifier (PV) Assay Sheets.
Root cause
Under Investigation by firm
Action
On May 11, 2020, Ortho Clinical Diagnostics, Inc. sent an Urgent Medical Device Correction Notification to their customers alerting them of important information regarding the precision of VITROS AMYL Slides. The notice stated that the imprecision affects serum, urine and QC samples, however VITROS AMYL Slides performance continues to meet its intended clinical use. The VITROS Performance Verifiers I and II Control Assay Sheet estimated within-lab SD, Range of Means (ROM) and Delta table values were re-evaluated. The Delta tables will be updated in a future revision of the Performance Verifier Training Module. In the meantime, customers were instructed to use the revised estimated within-lab SD listed in the enclosed Information sheet to evaluate their lab's quality control when using VITROS Performance Verifier I and II products. When available on the Ortho website, https://www.orthoclinicaldiagnostics.com, customers can obtain the updated VITROS Performance Verifier Control Assay Sheets and can use the revised ROM to evaluate the labs VITROS AMYL Slide quality control when using VITROS Performance Verifier I and II products. The firm explained that a review of previously reported patient results is NOT recommended. The customers were asked to Complete the Confirmation of Receipt form no later than May 19, 2020 and to forward the notification if the product was distributed outside of their facility.
Quantity
73,071
Distribution
Worldwide distribution - US Nationwide including in the statesof Maine, Massachusetts, Vermont, Connecticut, New Jersey, New York, Pennsylvania, Washington DC, Maryland, Virginia, West Virginia, North Carolina, South Carolina, Georgia, Florida, Alabama, Tennessee, Missouri, Mississippi, Kentucky, Ohio, Indiana, Michigan, Iowa, Wisconsin, Minnesota, South Dakota, North Dakota, Montana, Illinois, Kansas, Nebraska, Louisiana, Arkansas, Oklahoma, Texas, Colorado, Idaho, Utah, Arizona, New Mexico, California, Hawaii, Oregon , Washington, and Alaska
Lot, serial or model codes
Lot #s 600203564949, 600203564950, 600203564975, 600203564976, 600203564977, 600203564978, 600203566275, 600203566276, 600203566277, 600203566537, 600203566538, 600203566620, 600203566621, 600203566622, 600203566698, 600203566699, 600203566700, 600203567485, 600203567486, 600203567546, 600203567547, 600303570416, 600303570417, 600303570476, 600303570495, 600303570496, 600303571564, 600303571565, 600303576365, 600303576366, 600303578980, 600303578981, 600303578982, 600303578984, 600303578985, 600303579304, 600303579305, 600303579338, 600303579339, 600303579545, 600303579546, 600303579570, 600303579571, 600503581579, 620503581598, 600503581599, 600503581600, 600503581626, 600503582485, 600503582661, 600503582711, 600503582712, 600503582713, 600503583174, 600503583175, 600503584118, 600503584168, 600503584169, 600503589810, 600503589811, 600503589812, 600603591744, 600603591745, 600603591746, 600603595598, 600603595599, 600603595644, 600603595645, 600603595655, 600603595656, 600603597378, 600603597379, 600603599568, 600603599572, 600603599630, 600803637422, 600803637423, 600803637424, 600803638469, 600803638470
510(k) numbers
["K812027"]
PMA numbers
not on file
Product code
CIW
Firm FEI
1000305840
Firm city
Rochester
Firm state
NY
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register