Customed, Inc: LAPAROTOMY PACK- (1) MAYO STAND COVER REINFORCED LIF (10) GAUZE SPONGE 4" X 4" 16PLY XRD LIF (5) LAP SPONGES PRE-WASH 18" X 18" XRD LIF (1) LITE GLOVE LIF (1) BAG SUTURE FLORAL (1) BLADE SURGICAL #1 0 CARBON STEEL (1) BLADE SURGICAL #15 CARBON STEEL (1) YANKAUER SUCTION TUBE WITHOUT VENT LIF (1) CAUTERY TIP POLISHER LIF (1) BAG GLASSINE (1) TABLE COVER 44" X 90" L/F (2) DRAPE SHEET 41 " X 58" SMS (1) LAPAROTOMY DRAPE "T" 102" X 78" X 121" STD SMS (4) DRAPE UTILITY WITH TAPE LIF (2) NEEDLE & BLADE COUNTER 10C MAG/CLEAR LIF (1) TUBE SUCTION CONECT W' X 12' LIF (3) ABSORBENT TOWELS 15" X 20" LIF (1) CAUTERY PENCIL PUSH BOTTON (1) GOWN IMPERVIOUS EXTRA REINFORCED X-LARGE TOWEL/WRAP LEVEL Ill (2) GOWN X-LARGE SMS IMPERVIOUS REINFORCED LIF EO sterilized surgical convenient packs packaged into polyester/polyethylene breather bag with Tyvek double vents.

Recall number
Z-2364-2014
Recalling firm
Customed, Inc
Product
LAPAROTOMY PACK- (1) MAYO STAND COVER REINFORCED LIF (10) GAUZE SPONGE 4" X 4" 16PLY XRD LIF (5) LAP SPONGES PRE-WASH 18" X 18" XRD LIF (1) LITE GLOVE LIF (1) BAG SUTURE FLORAL (1) BLADE SURGICAL #1 0 CARBON STEEL (1) BLADE SURGICAL #15 CARBON STEEL (1) YANKAUER SUCTION TUBE WITHOUT VENT LIF (1) CAUTERY TIP POLISHER LIF (1) BAG GLASSINE (1) TABLE COVER 44" X 90" L/F (2) DRAPE SHEET 41 " X 58" SMS (1) LAPAROTOMY DRAPE "T" 102" X 78" X 121" STD SMS (4) DRAPE UTILITY WITH TAPE LIF (2) NEEDLE & BLADE COUNTER 10C MAG/CLEAR LIF (1) TUBE SUCTION CONECT W' X 12' LIF (3) ABSORBENT TOWELS 15" X 20" LIF (1) CAUTERY PENCIL PUSH BOTTON (1) GOWN IMPERVIOUS EXTRA REINFORCED X-LARGE TOWEL/WRAP LEVEL Ill (2) GOWN X-LARGE SMS IMPERVIOUS REINFORCED LIF EO sterilized surgical convenient packs packaged into polyester/polyethylene breather bag with Tyvek double vents.
Status
Terminated
Initiated
2014-05-20
Posted
2014-08-29
Terminated
2017-08-14
Reason
Customed has determined that there is the possibility that packaging integrity may be compromised on different catalog numbers (multiple lots). This could result in an injury to the patient due to product contamination or loss of sterility condition.
Root cause
Package design/selection
Action
Customed Inc. sent a Recall Notification letter dated June 3, 2014 via email and followed by letter to all affected customers. The letter identified the affected products, problem and actions to be taken. The letter informed customers if the affected products are in their inventory to: 1. Remove products from sale (quarantine them) and cease distribution. 2. Complete and return the enclosed acknowledgement form to Customed Inc .Recall Coordinator. Affected products should be returned to Customed Inc for credit. Customers with questions were instructed to contact Customed Inc. Recall Coordinator at 1-787-801-0100 Ext. 7540
Quantity
34 lots; 3603 units (multiple units per lot)
Distribution
Worldwide Distribution - US (nationwide) in the state of Florida including Puerto Rico (USA); & USVI (St John) and the countries of Tortola, Trinidad, Dominican Republic, Costa Rica and Mexico.
Lot, serial or model codes
Product code 900-996, 34 lots: 111040982 111051305 111061694 111071947 111082120 111092410 111092465 111102789 111113147 111123308 112010213 112020496 112030757 112041365 112052027 112072880 112083444 112093958 112114585 112125124 113025974 113036393 113057517 113057612 113057822 113078480 113078577 113099300 113099581 131110081 131110607 140111262 140211738 140513485
510(k) numbers
not on file
PMA numbers
not on file
Product code
FHI
Firm FEI
2648727
Firm city
Fajardo
Firm state
not on file
Firm country
PR

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register