SEASPINE ORTHOPEDICS CORPORATION: Brand Name: Meridian Angled Awl Fixed Product Name: Angled Awl Fixed Model/Catalog Number: RA2-003015 Product Description: Surgical Awl Component: No
- Recall number
- Z-2357-2024
- Recalling firm
- SEASPINE ORTHOPEDICS CORPORATION
- Product
- Brand Name: Meridian Angled Awl Fixed Product Name: Angled Awl Fixed Model/Catalog Number: RA2-003015 Product Description: Surgical Awl Component: No
- Status
- Open, Classified
- Initiated
- 2024-06-04
- Posted
- 2024-07-15
- Terminated
- not on file
- Reason
- Insufficient weld around the cap component of the instrument is insufficient to withstand impaction forces.
- Root cause
- Process design
- Action
- On June 4, 2024, SeaSpine Orthopedics issued a "Urgent Medical Device Recall" Notification via E-Mail to affected consignees. SeaSpine Orthopedics ask consignees to take the following actions: We request that you return the affected part to SeaSpine s Carlsbad facility for evaluation and potential rework. To support your upcoming cases, a replacement Vu aPod Prime NM Angled Awl will be sent (PN: 33-40-1028). Additionally, please continue to utilize the loaner pool. Please use the attached label to return the inventory in your possession at your earliest convenience. Once the inventory has been returned, please acknowledge via DocuSign.
- Quantity
- 100 units
- Distribution
- The product was distributed to the following US states: Nevada, Colorado, Texas, Washington, Indiana, Georgia, Michigan, Florida, Pennsylvania, Illinois, Tennessee, Virginia, Missouri, South Dakota, California, Montana, Minnesota, Arizona
- Lot, serial or model codes
- Lot Code: UDI 10889981216368 Lot Numbers TT0190H, TT0191H
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- HWJ
- Firm FEI
- 3012120772
- Firm city
- Carlsbad
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28