FDA Medical Device Recalls (openFDA)
Customed, Inc: ARTHROSCOPY PACK- (2) ABSORBENT TOWELS 15" x 20" LIF (1) TUBE SUCTION CONNECT Y.." X 12' LIF (1) ELASTIC BANDAGE 6" X 5yrd. UF (1) IMPERVIOUS STOCKINETTE 14" X 48" (1) HUCK CLOTH TOWEL WHITE (1) MAYO STAND COVER REINFORCED LIF (1) U DRAPE 60" X70" WITH TAPE SPLIT (4) STRIPS TAPE 24" X 4" L/F (1) STOCKINETTE 6" X 48" LIF (1) MAYO TRAY SMALL (5) HUCK CLOTH TOWELS BLUE (2) UTILITY BOWL 32oz (2) GOWN SURGICAL REINFORCED X-LARGE LEVEL IV (1) SOLUTION SURGICAL DURAPREP 26ml UF (1) SHEET EXTREMITY ABS. 125" x 85" SMS (5) LAP SPONGE PREWASH XRD (1) YANKAUER SUCTION TUBE WITHOUT VENT LIF (1) PENCIL CAUTERY ROCKER SWITCH (1) SKIN MARKER INK WITH 8 LABEL (2) LITES GLOVES UF (2) NEEDLE & BLADE COUNTER 10C MAG/CLEAR (1) CAUTERY TIP POLISHER UF (4) DRAPE UTILITY WITH TAPE UF (1) BAG BIOHAZARD 30" X 37" (1) BAG SUTURE FLORAL (1) TABLE COVER BTC 77" X 110" HD FULL COVERAGE UF (1) GOWN IMPERVIOUS EXTRA REINFORCED X-LARGE TOWEL/WRAP EO sterilized surgical convenient packs packaged into polyester/polyethylene breather bag with Tyvek double vents.
- Recall number
- Z-2356-2014
- Recalling firm
- Customed, Inc
- Product
- ARTHROSCOPY PACK- (2) ABSORBENT TOWELS 15" x 20" LIF (1) TUBE SUCTION CONNECT Y.." X 12' LIF (1) ELASTIC BANDAGE 6" X 5yrd. UF (1) IMPERVIOUS STOCKINETTE 14" X 48" (1) HUCK CLOTH TOWEL WHITE (1) MAYO STAND COVER REINFORCED LIF (1) U DRAPE 60" X70" WITH TAPE SPLIT (4) STRIPS TAPE 24" X 4" L/F (1) STOCKINETTE 6" X 48" LIF (1) MAYO TRAY SMALL (5) HUCK CLOTH TOWELS BLUE (2) UTILITY BOWL 32oz (2) GOWN SURGICAL REINFORCED X-LARGE LEVEL IV (1) SOLUTION SURGICAL DURAPREP 26ml UF (1) SHEET EXTREMITY ABS. 125" x 85" SMS (5) LAP SPONGE PREWASH XRD (1) YANKAUER SUCTION TUBE WITHOUT VENT LIF (1) PENCIL CAUTERY ROCKER SWITCH (1) SKIN MARKER INK WITH 8 LABEL (2) LITES GLOVES UF (2) NEEDLE & BLADE COUNTER 10C MAG/CLEAR (1) CAUTERY TIP POLISHER UF (4) DRAPE UTILITY WITH TAPE UF (1) BAG BIOHAZARD 30" X 37" (1) BAG SUTURE FLORAL (1) TABLE COVER BTC 77" X 110" HD FULL COVERAGE UF (1) GOWN IMPERVIOUS EXTRA REINFORCED X-LARGE TOWEL/WRAP EO sterilized surgical convenient packs packaged into polyester/polyethylene breather bag with Tyvek double vents.
- Status
- Terminated
- Initiated
- 2014-05-20
- Posted
- 2014-08-29
- Terminated
- 2017-08-14
- Reason
- Customed has determined that there is the possibility that packaging integrity may be compromised on different catalog numbers (multiple lots). This could result in an injury to the patient due to product contamination or loss of sterility condition.
- Root cause
- Package design/selection
- Action
- Customed Inc. sent a Recall Notification letter dated June 3, 2014 via email and followed by letter to all affected customers. The letter identified the affected products, problem and actions to be taken. The letter informed customers if the affected products are in their inventory to: 1. Remove products from sale (quarantine them) and cease distribution. 2. Complete and return the enclosed acknowledgement form to Customed Inc .Recall Coordinator. Affected products should be returned to Customed Inc for credit. Customers with questions were instructed to contact Customed Inc. Recall Coordinator at 1-787-801-0100 Ext. 7540
- Quantity
- 11 lots; 273 units (multiple units per lot)
- Distribution
- Worldwide Distribution - US (nationwide) in the state of Florida including Puerto Rico (USA); & USVI (St John) and the countries of Tortola, Trinidad, Dominican Republic, Costa Rica and Mexico.
- Lot, serial or model codes
- Product code 900-902, 11 lots: 112041360 112051755 112052021 113015521 113025983 113036389 113067942 113078352 140312189 140412917 140513342
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- FWZ
- Firm FEI
- 2648727
- Firm city
- Fajardo
- Firm state
- not on file
- Firm country
- PR
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28