Skip to the content

FDA Medical Device Recalls (openFDA)

Acuity Surgical Devices, LLC: A LINK Z INTERVERTEBRAL BODY FUSION DEVICE Product Description: A Link Z Ti Unitary Open Interbody 32x24x12 7 A Link Z Ti Unitary Open Interbody 32x24x12 12 A Link Z Ti Unitary Open Interbody 32x24x16 7 A Link Z Ti Unitary Open Interbody 32x24x16 12 Product Usage: is a spinal device that is implanted in the intervertebral body space via an anterior approach to improve stability of the spine while supporting fusion.

Recall number
Z-2355-2020
Recalling firm
Acuity Surgical Devices, LLC
Product
A LINK Z INTERVERTEBRAL BODY FUSION DEVICE Product Description: A Link Z Ti Unitary Open Interbody 32x24x12 7 A Link Z Ti Unitary Open Interbody 32x24x12 12 A Link Z Ti Unitary Open Interbody 32x24x16 7 A Link Z Ti Unitary Open Interbody 32x24x16 12 Product Usage: is a spinal device that is implanted in the intervertebral body space via an anterior approach to improve stability of the spine while supporting fusion.
Status
Open, Classified
Initiated
2019-01-07
Posted
not on file
Terminated
not on file
Reason
Due to a manufacturing process issue, it allowed vertebral body bone screws to not seat properly within the spinal device/cage.
Root cause
Process control
Action
On 12/14/2018, the firm notified its supplier via email requesting to visually review its inventory which determined that some screw holes were oval shaped. On 7/23/2019, all inventory was quarantined as part of the removal and was scrapped from inventory. Due to the low risk to patients and low probability, the firm ownly initially begin removal of product from internal locations (main warehouse) for the purpose of inspection and if necessary, repair or destruction. Removal Summary No additional customer or distributor communication took place during the removal.
Quantity
A total of 27 lots producing 178 units
Distribution
US Nationwide distribution including in the states of TX, FL, NC, LA, KY, WI, SD, IL, CA, and KS. O.U.S.: None
Lot, serial or model codes
Product Product Description Lot Number Number 15-32241207 A Link Z Ti Unitary Open Interbody 32x24x12 7 1000572 15-32241212 A Link Z Ti Unitary Open Interbody 32x24x12 12 1000573 15-32241607 A Link Z Ti Unitary Open Interbody 32x24x16 7 1000576 15-32241612 A Link Z Ti Unitary Open Interbody 32x24x16 12 1000577
510(k) numbers
["K133827"]
PMA numbers
not on file
Product code
OVD
Firm FEI
3010866843
Firm city
Irving
Firm state
TX
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register