Skip to the content

FDA Medical Device Recalls (openFDA)

Terumo Cardiovascular Systems Corporation: Terumo Advanced Perfusion System 1 Electronic Patient Gas System Catalog #: 801188

Recall number
Z-2350-2021
Recalling firm
Terumo Cardiovascular Systems Corporation
Product
Terumo Advanced Perfusion System 1 Electronic Patient Gas System Catalog #: 801188
Status
Terminated
Initiated
2021-06-08
Posted
not on file
Terminated
2022-06-28
Reason
intermittent failure causes a blender initialization fault upon start-up, or during an EPGS calibration attempt following the 15-minute warm-up period. The control sliders on the Central Control Monitor (CCM) are disabled and the gas system can only be controlled by the local control knobs
Root cause
Nonconforming Material/Component
Action
Terumo notified the regional affiliate to address the single EPGS device in Panama via email on 6/8/21. Letter states reason for recall, health risk and action to take: Terumo CVS will execute a field correction to replace the one (1) affected EPGS device. A Terumo Representative will contact you to schedule the field correction activities. 1. Review this Medical Device Correction and assure the user has received notice of this issue. 2. A Terumo representative will contact the user to schedule the field correction activity. Note: Terumo CVS recommends that users continue using the affected Terumo System 1 while waiting for this correction. We encourage you to contact your local Terumo representative with any questions or concerns.
Quantity
1 unit OUS
Distribution
International distribution in the country of Panama.
Lot, serial or model codes
S/N: 01751 UDI: 00886799000588
510(k) numbers
["K172220"]
PMA numbers
not on file
Product code
DTQ
Firm FEI
1828100
Firm city
Ann Arbor
Firm state
MI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register