FDA Medical Device Recalls (openFDA)
Terumo Cardiovascular Systems Corporation: Terumo Advanced Perfusion System 1 Electronic Patient Gas System Catalog #: 801188
- Recall number
- Z-2350-2021
- Recalling firm
- Terumo Cardiovascular Systems Corporation
- Product
- Terumo Advanced Perfusion System 1 Electronic Patient Gas System Catalog #: 801188
- Status
- Terminated
- Initiated
- 2021-06-08
- Posted
- not on file
- Terminated
- 2022-06-28
- Reason
- intermittent failure causes a blender initialization fault upon start-up, or during an EPGS calibration attempt following the 15-minute warm-up period. The control sliders on the Central Control Monitor (CCM) are disabled and the gas system can only be controlled by the local control knobs
- Root cause
- Nonconforming Material/Component
- Action
- Terumo notified the regional affiliate to address the single EPGS device in Panama via email on 6/8/21. Letter states reason for recall, health risk and action to take: Terumo CVS will execute a field correction to replace the one (1) affected EPGS device. A Terumo Representative will contact you to schedule the field correction activities. 1. Review this Medical Device Correction and assure the user has received notice of this issue. 2. A Terumo representative will contact the user to schedule the field correction activity. Note: Terumo CVS recommends that users continue using the affected Terumo System 1 while waiting for this correction. We encourage you to contact your local Terumo representative with any questions or concerns.
- Quantity
- 1 unit OUS
- Distribution
- International distribution in the country of Panama.
- Lot, serial or model codes
- S/N: 01751 UDI: 00886799000588
- 510(k) numbers
- ["K172220"]
- PMA numbers
- not on file
- Product code
- DTQ
- Firm FEI
- 1828100
- Firm city
- Ann Arbor
- Firm state
- MI
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28