Skip to the content

FDA Medical Device Recalls (openFDA)

Stryker Howmedica Osteonics Corp.: Rejuvenate Broach Catalog No: 1601-1005, 1601-1006, 1601-1007, 1601-1008, 1601-1009, 1601-1010 Stryker Orthopaedics, Mahwah, NJ. 07430 Contours the proximal femur to the geometry of the stem to provide a complimentary fit of the stem to the bone.The broaches are designed with a series of teeth that are intended to seat against highly densified cancellous bone for the interference press-fit in the uncemented application.

Recall number
Z-2349-2008
Recalling firm
Stryker Howmedica Osteonics Corp.
Product
Rejuvenate Broach Catalog No: 1601-1005, 1601-1006, 1601-1007, 1601-1008, 1601-1009, 1601-1010 Stryker Orthopaedics, Mahwah, NJ. 07430 Contours the proximal femur to the geometry of the stem to provide a complimentary fit of the stem to the bone.The broaches are designed with a series of teeth that are intended to seat against highly densified cancellous bone for the interference press-fit in the uncemented application.
Status
Terminated
Initiated
2008-01-25
Posted
2008-09-19
Terminated
2008-09-22
Reason
Rejuvenate Broach system was designed in a manner that may result in the potential for a medical calcar fracture.
Root cause
Device Design
Action
Stryker notified all branches and agencies to return the product on January 25, 2008. A formal recall letter was not written as all product was in their control.
Quantity
78 devices
Distribution
Nationwide Distribution.
Lot, serial or model codes
All Lot Codes
510(k) numbers
not on file
PMA numbers
not on file
Product code
HTQ
Firm FEI
3003070421
Firm city
Mahwah
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register