Skip to the content

FDA Medical Device Recalls (openFDA)

Kyphon Inc: KyphX Osteo Introducer System with blunt tip introducer stylet (as contained in the KyphoPak Tray, First Fracture, 15/3 Xpander), catalog number KPT1502, manufactured by Medtronic Spine LLC. Device is used to access vertebral body during balloon kyphoplasty, a procedure designed to reduce and stabilize vertebral body compression fractures (VCFS).

Recall number
Z-2347-2008
Recalling firm
Kyphon Inc
Product
KyphX Osteo Introducer System with blunt tip introducer stylet (as contained in the KyphoPak Tray, First Fracture, 15/3 Xpander), catalog number KPT1502, manufactured by Medtronic Spine LLC. Device is used to access vertebral body during balloon kyphoplasty, a procedure designed to reduce and stabilize vertebral body compression fractures (VCFS).
Status
Terminated
Initiated
2008-06-23
Posted
2008-09-18
Terminated
2009-10-07
Reason
Some products may contain the incorrect introducer stylet. If the treating physician is unfamiliar with the use and design of the diamond tip stylet, there is a possibility of serious injuring occurring with the use of the diamond tip stylet during balloon kyphoplasty procedure. Other risks include delay to surgery and inconvenience to patient.
Root cause
Mixed-up of materials/components
Action
Recall initiated on June 23, 2008. A recall letter and reconciliation form was sent out via e-mail to sales representatives for hand delivery to US hospital consignees. The consignees were asked to review the letter and complete/return the reconciliation form to the firm. If you have any questions contact Kyphon, Inc. at 1-408-548-6740.
Quantity
310 units
Distribution
Class II Recall - Nationwide Distribution.
Lot, serial or model codes
Lot code J8052810
510(k) numbers
not on file
PMA numbers
not on file
Product code
LXH
Firm FEI
3004904811
Firm city
Sunnyvale
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register