Dynex Technologies, Inc.: Dynex Agility Sample Tips; Model Number: 67910; The 67910 Agility Sample Tips are an accessory/consumable used with the Agility Automated ELISA system.

Recall number
Z-2345-2025
Recalling firm
Dynex Technologies, Inc.
Product
Dynex Agility Sample Tips; Model Number: 67910; The 67910 Agility Sample Tips are an accessory/consumable used with the Agility Automated ELISA system.
Status
Open, Classified
Initiated
2025-07-16
Posted
2025-08-22
Terminated
not on file
Reason
The internal label affixed to the sample tip rack is incorrectly labeled as reagent tip racks. When scanned by the Agility instrument, the tip rack is misidentified as a different tip type, rendering it incompatible with the system and preventing use which causes a delay in obtaining patient results.
Root cause
Labeling Change Control
Action
On July 16, 2025, URGENT FIELD SAFETY NOTICE letters were emailed to customers. Action to Be Taken by the User: - Upon receipt, Quarantine affected devices and return or destroy them. - Customers have been provided a Field Safety Notice Customer Reply Form. Please complete this form, sign it and return to Dynex as soon as possible to acknowledge that you have received this information and have destroyed or returned to Dynex the impacted 67910 Agility Sample Tips Lot 863366. Transmission of this Field Safety Notice This notice needs to be passed on all those who need to be aware within your organisation or to any organisation where the potentially affected devices have been transferred. (As appropriate) Please transfer this notice to other organisations on which this action has an impact. (As appropriate) Please maintain awareness on this notice and resulting action for an appropriate period to ensure effectiveness of the corrective action. Please report all device-related incidents to the manufacturer, distributor or local representative, and the national Competent Authority if appropriate, as this provides important feedback..*
Quantity
237 units
Distribution
Domestic: CA, FL, NC, NJ, NY;
Lot, serial or model codes
Model Number: 67910; UDI-DI: 5060456180072; Lot Number: 863366;
510(k) numbers
not on file
PMA numbers
not on file
Product code
LXG
Firm FEI
1117676
Firm city
Chantilly
Firm state
VA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this source