Fresenius Medical Care Holdings, Inc.: RTL190828: 2008K@HOME HEMODIALYSIS SYSTEM - Blood Pump Rotor spare part, Model Number: F40015481 Rev A

Recall number
Z-2345-2024
Recalling firm
Fresenius Medical Care Holdings, Inc.
Product
RTL190828: 2008K@HOME HEMODIALYSIS SYSTEM - Blood Pump Rotor spare part, Model Number: F40015481 Rev A
Status
Open, Classified
Initiated
2024-06-03
Posted
2024-07-11
Terminated
not on file
Reason
An increased number of complaints were reported on the Blood Pump Rotor (P/N - F40015481 REV A) after a period of use by clinics noting loose/dislodged guide sheaves.
Root cause
Component design/selection
Action
On June 3, 2024 IMPORTANT MEDICAL DEVICE CORRECTION letters were sent to consignees. Fresenius Medical Care is requesting replacement of affected Blood Pump Rotors in 2008K@HOME systems due to the patients possibly being unaware of this component failing. To order for a replacement Blood Pump Rotor (Part Number, F40015481), contact Fresenius Medical Care Customer Service at 800-227-2572, OPTION 3, o Please have Blood Pump Rotor Part Number and Lot Number available when requesting a replacement. CUSTOMER: 1. Please provide this letter to your Biomedical Technician. 2. Report any complaints or adverse events to product.complaints@fmc-na.com or FDA MedWatch at https://www.fda.gov/safety/medical-product-safety-information/medwatch-forms-fda-safety-reporting. 3. Provide this Customer Notice to all those who need to be aware within your organization. 4. Complete and return the attached reply form. For any clinical or medical questions, please contact the Global Medical Information and Education Office at 1-855-616-2309.
Quantity
1 unit
Distribution
Domestic: Nationwide Distribution. International: Canada.
Lot, serial or model codes
Model Number: RTL190828 (Rental Units); UDI/DI: 00840861100941; Serial Number: 1K0S172181
510(k) numbers
not on file
PMA numbers
not on file
Product code
KDI
Firm FEI
3001451489
Firm city
Waltham
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register