FDA Medical Device Recalls (openFDA)
Siemens Medical Solutions USA, Inc: MAGNETOM Prisma magnetic resonance diagnostic devices Model #10849582
- Recall number
- Z-2345-2019
- Recalling firm
- Siemens Medical Solutions USA, Inc
- Product
- MAGNETOM Prisma magnetic resonance diagnostic devices Model #10849582
- Status
- Terminated
- Initiated
- 2019-07-09
- Posted
- not on file
- Terminated
- 2020-04-23
- Reason
- Potential of an open port in the quench venting system when using the optional horizontal outlet on the quench line assembly.
- Root cause
- Device Design
- Action
- 1. review the System Owner Manual chapter 1.3.3 Quench emergency plan and the section regarding defective vent line. 2. follow the standard emergency procedures you have in place in the event of system failures. 3.A Siemens Service Technician will conduct an inspection to verify that your system has a vertical outlet. If a vertical outlet is present, no further action is required. In the event your system is equipped with a horizontal outlet, the Siemens Service Technician will ensure that the port for the drop-off plate is adequately blanked off or sealed. 4.service organization will contact you shortly to arrange a date to perform this corrective action. Please feel free to contact our service organization for an earlier appointment at 1-800-888-7436. 5. Please review, understand, and confirm acknowledgement of the information and instructions outlined in this Customer Safety Advisory Notice. By signing this document you are acknowledging that you have read and understand the content therein. If you have any question, please contact Meredith Adams, Telephone 610-448-6461 and E-Mail: Meredith.adams@siemens-healthineers.com
- Quantity
- 17
- Distribution
- AL AR AZ CA CO CT DC DE FL GA GU IA ID IL IN KS KY LA MA MD ME MI MN MO MT NC NE NH NJ NY OH OK OR PA SC TN TX VA WA WI WV WY
- Lot, serial or model codes
- Serial # 66109 66111 166049 66047 166037 66102 66027 66053 66094 66028 66069 66092 66046 66082 166097 66044 166082
- 510(k) numbers
- ["K132119"]
- PMA numbers
- not on file
- Product code
- LNH
- Firm FEI
- 2240869
- Firm city
- Malvern
- Firm state
- PA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28