Siemens Medical Solutions USA, Inc: Artis zee/ zeego, Artis Q/ Q.zen, stand alone system, software controlled Model numbers: 10094135, 10094137, 10094139, 10094141, 10280959, 10848281, 10848282, 10848283, 10848353, 10848255 Artis zee / zeego and Q/ Q.zen is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures that can be performed with the Artis zee / zeego and Q/ Q.zen include cardiac angiography, neuro angiography, general angiography, rotational angiography, operating room angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures. Artis zee / zeego and Q/ Q.zen can also support the acquisition of position triggered imaging for spatial data synthesis.

Recall number
Z-2344-2016
Recalling firm
Siemens Medical Solutions USA, Inc
Product
Artis zee/ zeego, Artis Q/ Q.zen, stand alone system, software controlled Model numbers: 10094135, 10094137, 10094139, 10094141, 10280959, 10848281, 10848282, 10848283, 10848353, 10848255 Artis zee / zeego and Q/ Q.zen is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures that can be performed with the Artis zee / zeego and Q/ Q.zen include cardiac angiography, neuro angiography, general angiography, rotational angiography, operating room angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures. Artis zee / zeego and Q/ Q.zen can also support the acquisition of position triggered imaging for spatial data synthesis.
Status
Terminated
Initiated
2016-07-07
Posted
not on file
Terminated
2016-12-13
Reason
Due to an error in the 19 Live Display, image reproduction may fail in the examination room and the potential exists for the loss of images immediately after system startup.
Root cause
Nonconforming Material/Component
Action
Siemens sent an Important Customer Safety Notice dated July 7, 2016, by certified mail to customers to inform them of the corrective action they plan to take to prevent potential loss of the image display for live images in the examination room after system startup. Siemens service organization will contact each customer to arrange a date to resolve the issue with a software update of the monitor firmware via AX012/16/S. For questions regarding this recall call 610-448-6461.
Quantity
51 units
Distribution
Nationwide Distribution to AL,AZ,CA,CO,IA,IL,MA,MD,MI,MN,MO,NE,NJ,NY,OH,OR,PA,RI,TX,VA,WA,WI,WV
Lot, serial or model codes
Serial Numbers: 121185,147996,121169,105032,109190,109191,109197,109198,125100,158225,161021,154847,124083,123050,109187,124102,121167,125021,121163,117125,148001,158234,148010,158229,121171, 109183,109181,131101,158226,109194,117124,158244,109179,158238,109188,121179,137714,121168,158246,161022,121175,121173,154859,109186,158230,123046,158227,109189,123052,105029,105030
510(k) numbers
["K073290","K123529"]
PMA numbers
not on file
Product code
IZI
Firm FEI
2240869
Firm city
Malvern
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register