Skip to the content

FDA Medical Device Recalls (openFDA)

Avanos Medical, Inc.: Product Name: MIC Safety Percutaneous Endoscopic Gastrostomy PEG Kit 24 Fr PUSH OTW Model/Catalog Number: 7170-24 Product Description: Feeding Tube Kit Component: No

Recall number
Z-2343-2026
Recalling firm
Avanos Medical, Inc.
Product
Product Name: MIC Safety Percutaneous Endoscopic Gastrostomy PEG Kit 24 Fr PUSH OTW Model/Catalog Number: 7170-24 Product Description: Feeding Tube Kit Component: No
Status
Open, Classified
Initiated
2026-04-30
Posted
2026-06-05
Terminated
not on file
Reason
Lidocaine hydrochloride injection, the subject of a supplier recall for quality issues, were included into sterile kits and sets.
Root cause
Material/Component Contamination
Action
On May 1, 2026, Avanaos Medical, Inc. issued a "Urgent: Medical Device Recall" Notification to affected consignees via Email. Avanos asked consignees to take the following actions: 1. Immediately review your inventory and identify any affected product. 1a. Quarantine all affected lots to prevent further use or distribution. 1b. Discontinue use of the affected lidocaine hydrochloride injection 1%, 5 mL and return the entire unused kit to Avanos Medical, Inc. for destruction. 2. This notice needs to be passed on to all those who need to be aware within your organization or functions where the potentially affected devices have been transferred. 3. Complete and return the Customer Response Form as instructed, including identifying whether you have inventory of the affected product available for return. Distributors: 1. Provide customers a copy of this notification. Instruct customers to follow the actions outlined in this letter. 2. Provide refunds or credits to your customers in accordance with your internal processes. 3. Submit an updated Customer Response Form to Avanos to account for product distributed to your customers and to receive applicable replacement product.
Quantity
132 kits
Distribution
US Nationwide distribution.
Lot, serial or model codes
UDI-DI: 10350770007790, Lot Numbers 30370715, 30371329, 30375466
510(k) numbers
["K924065"]
PMA numbers
not on file
Product code
KNT
Firm FEI
3011270181
Firm city
Alpharetta
Firm state
GA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register